Which sterilization method is the one most widely used in the medical office?

  • Journal List
  • Elsevier Public Health Emergency Collection
  • PMC7099662

Mandell, Douglas, and Bennett's Principles and Practice of Infectious Diseases. 2015 : 3294–3309.e4.

Guest Editor (s): John E. Bennett, MD, MACP

Adjunct Professor of Medicine, Uniformed Services University of the Health Sciences, F. Edward Hebert School of Medicine, Bethesda, Maryland

Guest Editor (s): Raphael Dolin, MD

Maxwell Finland Professor of Medicine (Microbiology and Molecular Genetics), Harvard Medical School, Boston, Massachusetts

Attending Physician, Beth Israel Deaconess Medical Center, Boston, Massachusetts

Brigham and Women' s Hospital, Boston, Massachusetts

Guest Editor (s): Martin J. Blaser, MD

Muriel G. and George W. Singer Professor of Translational Medicine, Professor of Microbiology, Director, Human Microbiome Program, Departments of Medicine and Microbiology, New York University School of Medicine, Langone Medical Center, New York, New York

Each year in the United States there are approximately 53 million outpatient surgical procedures and 46 million inpatient surgical procedures.1 For example, there are at least 10 million gastrointestinal endoscopies per year.2 Each of these procedures involves contact by a medical device or surgical instrument with a patient's sterile tissue or mucous membranes. A major risk of all such procedures is the introduction of infection. Failure to properly disinfect or sterilize equipment carries not only the risk associated with breach of the host barriers but also the additional risk for person-to-person transmission (e.g., hepatitis B virus) and transmission of environmental pathogens (e.g., Clostridium difficile).

Achieving disinfection and sterilization through the use of disinfectants and sterilization practices is essential for ensuring that medical and surgical instruments do not transmit infectious pathogens to patients. Because it is unnecessary to sterilize all patient care items, health care policies must identify whether cleaning, disinfection, or sterilization is indicated based primarily on the item's intended use.

Multiple studies in many countries have documented lack of compliance with established guidelines for disinfection and sterilization.3, 4 Failure to comply with scientifically based guidelines has led to numerous outbreaks of infectious diseases.2, 4, 5, 6, 7, 8 In this chapter, which is an update of previous chapters,9, 10, 11, 12, 13 a pragmatic approach to the judicious selection and proper use of disinfection and sterilization processes is presented, based on well-designed studies assessing the efficacy (via laboratory investigations) and effectiveness (via clinical studies) of disinfection and sterilization procedures. In addition, we briefly review the management of medical waste in health care facilities.

Definition of Terms

Sterilization is the complete elimination or destruction of all forms of microbial life and is accomplished in health care facilities by either physical or chemical processes. Steam under pressure, dry heat, ethylene oxide (ETO) gas, hydrogen peroxide gas plasma, vaporized hydrogen peroxide, and liquid chemicals are the principal sterilizing agents used in health care facilities. Sterilization is intended to convey an absolute meaning, not a relative one. Unfortunately, some health care professionals as well as the technical and commercial literature refer to “disinfection” as “sterilization” and items as “partially sterile.” When chemicals are used for the purposes of destroying all forms of microbiologic life, including fungal and bacterial spores, they may be called chemical sterilants. These same germicides used for shorter exposure periods may also be part of the disinfection process (i.e., high-level disinfection).

Disinfection describes a process that eliminates many or all pathogenic microorganisms on inanimate objects, with the exception of bacterial spores. Disinfection is usually accomplished by the use of liquid chemicals or wet pasteurization in health care settings. The efficacy of disinfection is affected by a number of factors, each of which may nullify or limit the efficacy of the process. Some of the factors that affect both disinfection and sterilization efficacy are the prior cleaning of the object; the organic and inorganic load present; the type and level of microbial contamination; the concentration of and exposure time to the germicide; the nature of the object (e.g., crevices, hinges, and lumens); the presence of biofilms; the temperature and pH of the disinfection process; and, in some cases, the relative humidity of the sterilization process (e.g., with ETO).

By definition then, disinfection differs from sterilization by its lack of sporicidal property, but this is an oversimplification. A few disinfectants will kill spores with prolonged exposure times (e.g., 3 to 12 hours) and are called chemical sterilants. At similar concentrations but with shorter exposure periods (e.g., 12 minutes for 0.55% ortho-phthalaldehyde) these same disinfectants will kill all microorganisms with the exception of large numbers of bacterial spores and are called high-level disinfectants. Low-level disinfectants may kill most vegetative bacteria, some fungi, and some viruses in a practical period of time (≤10 minutes), whereas intermediate-level disinfectants may be cidal for mycobacteria, vegetative bacteria, most viruses, and most fungi but do not necessarily kill bacterial spores. The germicides differ markedly among themselves primarily in their antimicrobial spectrum and rapidity of action.

Cleaning, on the other hand, is the removal of visible soil (e.g., organic and inorganic material) from objects and surfaces, and it normally is accomplished by manual or mechanical means using water with detergents or enzymatic products. Thorough cleaning is essential before high-level disinfection and sterilization because inorganic and organic materials that remain on the surfaces of instruments interfere with the effectiveness of these processes. Also, if the soiled materials become dried or baked onto the instruments, the removal process becomes more difficult and the disinfection or sterilization process less effective or ineffective. Surgical instruments should be presoaked or rinsed to prevent drying of blood and to soften or remove blood from the instruments. Decontamination is a procedure that removes pathogenic microorganisms from objects so they are safe to handle, use, or discard.

Terms with a suffix “-cide” or “-cidal” for killing action also are commonly used. For example, a germicide is an agent that can kill microorganisms, particularly pathogenic organisms (“germs”). The term germicide includes both antiseptics and disinfectants. Antiseptics are germicides applied to living tissue and skin, whereas disinfectants are antimicrobial agents applied only to inanimate objects. Preservatives are agents that inhibit the growth of microorganisms capable of causing biologic deterioration of substances/materials. In general, antiseptics are only used on the skin and not for surface disinfection and disinfectants are rarely used for skin antisepsis because they may cause injury to skin and other tissues. Other words with the suffix “-cide” (e.g., virucide, fungicide, bactericide, sporicide, and tuberculocide) can kill the type of microorganism identified by the prefix. For example, a bactericide is an agent that kills bacteria.14, 15, 16, 17, 18, 19

Rational Approach to Disinfection and Sterilization

About 45 years ago, Earle H. Spaulding15 devised a rational approach to disinfection and sterilization of patient care items or equipment. This classification scheme is so clear and logical that it has been retained, refined, and successfully used by infection control professionals and others when planning methods for disinfection or sterilization.* Spaulding believed that the nature of disinfection could be understood more readily if instruments and items for patient care were divided into three categories based on the degree of risk for infection involved in the use of the items. Although the scheme remains valid, some examples of disinfection studies with viruses, mycobacteria, and protozoa challenge the current definitions and expectations of high- and low-level disinfection.22 The three categories Spaulding described were critical, semicritical, and noncritical.

Critical Items

Critical items are so called because of the high risk for infection if such an item is contaminated with any microorganism, including bacterial spores. Thus, it is critical that objects that enter sterile tissue or the vascular system be sterile because any microbial contamination could result in disease transmission. This category includes surgical instruments, cardiac and urinary catheters, implants, arthroscopes, laparoscopes, and ultrasound probes used in sterile body cavities. Most of the items in this category should be purchased in sterile form or be sterilized by steam sterilization if possible. If heat sensitive, the object may be treated with ETO, hydrogen peroxide gas plasma, hydrogen peroxide vapor, or liquid chemical sterilants if other methods are unsuitable. Tables 301-1 and 301-2 list several germicides categorized as chemical sterilants and high-level disinfectants. These include 2.4% or greater glutaraldehyde-based formulations, hypochlorous acid/hypochlorite 650 to 675 ppm free chlorine, 1.12% glutaraldehyde with 1.93% phenol/phenate, 3.4% glutaraldehyde with 26% isopropanol,23 7.5% stabilized hydrogen peroxide, 2.0% hydrogen peroxide, 7.35% hydrogen peroxide with 0.23% peracetic acid, 8.3% hydrogen peroxide with 7.0% peracetic acid, 0.2% peracetic acid, 0.55% or greater ortho-phthalaldehyde, and 0.08% peracetic acid with 1.0% hydrogen peroxide.24 Liquid chemical sterilants can be relied on to produce sterility only if cleaning (to eliminate organic and inorganic material) precedes treatment and if proper use as to concentration, contact time, temperature, and pH is met.25

TABLE 301-1

Methods of Sterilization and Disinfection

STERILIZATION
DISINFECTION
Critical Items (will enter tissue or vascular system or blood will flow through them)
High-Level (semicritical items [except dental] will come in contact with mucous membrane or nonintact skin)
Intermediate-Level (some semicritical items1 and noncritical items)
Low-Level (noncritical items; will come in contact with intact skin)
ObjectProcedureExposure TimeProcedure (exposure time 12-45 min at ≥20° C2, 3)Procedure (exposure time ≥1 min9)Procedure (exposure time ≥1 min9)
Smooth, hard surface1, 4 A MR D
A MR E L5 L
C MR F M M
D 10 hr at 20-25° C G N N
F 6 hr H P O
G 12 min at 50°-56° C I6 Q P
J Q
H 3-8 hr K
Rubber tubing and catheters3, 4 A MR D
B MR E
C MR F
D 10 hr at 20°-25° C G
F 6 hr H
G 12 min at 50°-56° C I6
H 3-8 hr J
K
Polyethylene tubing and catheters3, 4, 7 A MR D
B MR E
C MR F
D 10 hr at 20°-25° C G
F 6 hr H
G 12 min at 50°-56° C I6
H 3-8 hr J
K
Lensed instruments4 A MR D
B MR E
C MR F
D 10 hr at 20°-25° C G
F 6 hr H
G 12 min at 50°-56° C J
H 3-8 hr K
Thermometers (oral and rectal)8 P8
Hinged instruments4 A MR D
B MR E
C MR F
D 10 hr at 20°-25° C G
F 6 hr H
G 12 min at 50°-56° C I6
H 3-8 hr J
K

Modified from the works of Rutala and Simmons and their colleagues.9, 10, 13, 16, 18, 19, 303

TABLE 301-2

Summary of Advantages and Disadvantages of Chemical Agents Used as Chemical Sterilants or as High-Level Disinfectants

STERILANT OR DISINFECTANTADVANTAGESDISADVANTAGES
Peracetic acid/hydrogen peroxide No activation required
Irritation not significant
Material compatibility concerns (lead, brass, copper, zinc) both cosmetic and functional
Limited clinical experience
Potential for eye and skin damage
Glutaraldehyde Numerous use studies published
Relatively inexpensive
Excellent material compatibility
Respiratory irritation from glutaraldehyde vapor
Pungent and irritating odor
Relatively slow mycobactericidal activity (unless other disinfectants added such as phenolic, alcohol)
Coagulates blood and fixes tissue to surfaces
Allergic contact dermatitis
Hydrogen peroxide, standard No activation required
May enhance removal of organic matter and organisms
No disposal issues
No odor or irritation issues
Does not coagulate blood or fix tissues to surfaces
Inactivates Cryptosporidium at high concentrations (e.g., 7.5%)
Use studies published
Material compatibility concerns (brass, zinc, copper, and nickel/silver plating) both cosmetic and functional
Serious eye damage with contact
Some studies show limited bactericidal activity of standard 3%
Ortho-phthalaldehyde Fast-acting high-level disinfectant
No activation required
Odor not significant
Excellent materials compatibility claimed
Efficacy data published
Does not coagulate blood or fix tissues to surfaces claimed
Stains protein gray (e.g., skin, mucous membranes, clothing, and environmental surfaces)
More expensive than glutaraldehyde
Eye irritation with contact
Slow sporicidal activity
Contraindicated for urologic instruments due to anaphylaxis
Peracetic acid Rapid cycle time (30-45 min)
Elevated temperature (50°-55° C) liquid immersion
Environmental friendly by-products (acetic acid, O2, H2O)
Fully automated endoscope reprocessing system
Single-use system eliminates need for concentration testing
Standardized cycle
May enhance removal of organic material and endotoxin
No adverse health effects to operators under normal operating conditions
Compatible with many materials and instruments
Does not coagulate blood or fix tissues to surfaces
Sterilant flows through scope facilitating salt, protein, and microbe removal
Rapidly sporicidal
Provides procedure standardization (constant dilution, perfusion of channel, temperatures, exposure)
Potential material incompatibility (e.g., aluminum anodized coating becomes dull)
Used for immersible instruments only
One scope or a small number of instruments can be processed in a cycle
More expensive (endoscope repairs, operating costs, purchase costs) than high-level disinfection
Serious eye and skin damage (concentrated solution) with contact
Point-of-use system, no long-term storage
Improved hydrogen peroxide (≥2.0%) No activation required
No odor
Nonstaining
No special venting requirements
Manual or automated applications
12-month shelf life, 14-day reuse
8 min at 20° C high-level disinfectant claim
Material compatibility concerns due to limited clinical experience
Organic material resistance concerns due to limited data
Limited clinical use and comparative microbicidal efficacy data
No measurable activity against Clostridium difficile spores

Modified from references 13, 93, 278, 304.

Semicritical Items

Semicritical items are those that come in contact with mucous membranes or nonintact skin. Respiratory therapy and anesthesia equipment, some endoscopes, laryngoscope blades and handles,26 esophageal manometry probes, endocavitary probes,26 nasopharyngoscopes, prostate biopsy probes,27 infrared coagulation device,28 anorectal manometry catheters, cystoscopes,29 and diaphragm fitting rings are included in this category.26 These medical devices should be free of all microorganisms, although small numbers of bacterial spores may be present. Intact mucous membranes, such as those of the lungs or the gastrointestinal tract, generally are resistant to infection by common bacterial spores but susceptible to other organisms such as bacteria, mycobacteria, and viruses. Semicritical items minimally require high-level disinfection using chemical disinfectants. Glutaraldehyde, hydrogen peroxide, ortho-phthalaldehyde, peracetic acid, and peracetic acid with hydrogen peroxide are cleared by the U.S. Food and Drug Administration (FDA) and are dependable high-level disinfectants provided the factors influencing germicidal procedures are met (see Tables 301-1 and 301-2). When a disinfectant is selected for use with certain patient care items, the chemical compatibility after extended use with the items to be disinfected also must be considered.

The complete elimination of all microorganisms in or on an instrument, with the exception of small numbers of bacterial spores, is the traditional definition of high-level disinfection. The FDA's definition of high-level disinfection is a sterilant used for a shorter contact time to achieve at least a 6-log10 kill of an appropriate Mycobacterium species. Cleaning followed by high-level disinfection should eliminate sufficient pathogens to prevent transmission of infection.30, 31

Semicritical items should be rinsed with sterile water after high-level disinfection to prevent their contamination with organisms that may be present in tap water, such as nontuberculous mycobacteria,8, 32 Legionella, 33, 34 or gram-negative bacilli such as Pseudomonas. 18, 20, 35, 36, 37 In circumstances where rinsing with sterile water rinse is not feasible, a tap water or filtered water (0.2-µm filter) rinse should be followed by an alcohol rinse and forced air drying.9, 37, 38, 39 Forced-air drying markedly reduces bacterial contamination of stored endoscopes, most likely by removing the wet environment favorable for bacterial growth.38 After rinsing, items should be dried and stored (e.g., packaged) in a manner that protects them from recontamination.

Some items that may come in contact with nonintact skin for a brief period of time (i.e., hydrotherapy tanks, bed side rails) are usually considered noncritical surfaces and are disinfected with low- or intermediate-level disinfectants (i.e., phenolic, iodophor, alcohol, chlorine).40 Because hydrotherapy tanks have been associated with spread of infection, some facilities have chosen to disinfect them with recommended levels of chlorine.40

Noncritical Items

Noncritical items are those that come in contact with intact skin but not mucous membranes. Intact skin acts as an effective barrier to most microorganisms; therefore, the sterility of items that come in contact with intact skin is “not critical.” Examples of noncritical items are bedpans, blood pressure cuffs, crutches, bed rails, bedside tables, patient furniture, and floors. The five most commonly touched noncritical items in the patient environment have been quantitatively shown to be bed rails, bed surface, supply cart, overbed table, and intravenous-line pump.41 In contrast to critical and some semicritical items, most noncritical reusable items may be decontaminated where they are used and do not need to be transported to a central processing area. There is virtually no documented risk of transmitting infectious agents to patients via noncritical items36 when they are used as noncritical items and do not contact nonintact skin or mucous membranes. However, these items (e.g., bedside tables, bed rails) could potentially contribute to secondary transmission by contaminating hands of health care workers or by contact with medical equipment that will subsequently come in contact with patients.14, 42, 43, 44, 45 Table 301-1 lists several low-level disinfectants that may be used for noncritical items. The exposure time listed in Table 301-1 is equal to or greater than 1 minute. Many U.S. Environmental Protection Agency (EPA)-registered disinfectants have a 10-minute label claim. However, multiple investigators have demonstrated the effectiveness of these disinfectants against vegetative bacteria (e.g., Listeria, Escherichia coli, Salmonella, vancomycin-resistant enterococci [VRE], methicillin-resistant Staphylococcus aureus [MRSA]), yeasts (e.g., Candida), mycobacteria (e.g., Mycobacterium tuberculosis), and viruses (e.g., poliovirus) at exposure times of 30 to 60 seconds.42, 43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54, 55, 56, 57, 58 Thus, it is acceptable to disinfect noncritical medical equipment (e.g., blood pressure cuff) and noncritical surfaces (e.g., bedside table) with an EPA-registered disinfectant or disinfectant/detergent at the proper use-dilution and a contact time of at least 1 minute.9, 59 Because the typical drying time for a germicide on a surface is 1 to 3 minutes (unless the product contains alcohol [e.g., a 60% to 70% alcohol will dry in about 30 seconds]) (N. Omidbakhsh, written communication), one application of the germicide on all hand contact surfaces to be disinfected is recommended.

Mops (microfiber and cotton string), reusable cleaning cloths, and disposable wipes are regularly used to achieve low-level disinfection.60, 61 Microfiber mops have demonstrated superior microbial removal compared with cotton string mops when used with detergent cleaner (95% vs. 68%, respectively). Use of a disinfectant did significantly improve microbial removal when a cotton string mop was used.61 Mops (especially cotton-string mops) are commonly not kept adequately cleaned and disinfected, and if the water-disinfectant mixture is not changed regularly (e.g., after every 3 to 4 rooms, no longer than 60-minute intervals), the mopping procedure may actually spread heavy microbial contamination throughout the health care facility.62 In one study, standard laundering provided acceptable decontamination of heavily contaminated mop heads but chemical disinfection with a phenolic was less effective.62 The frequent laundering of cotton-string mops (e.g., daily) is, therefore, recommended.

Hospital cleanliness continues to attract patient attention and in the United States it is still primarily assessed via visual appearance, which is not a reliable indicator of surface cleanliness.63 Three other methods have been offered for monitoring patient room hygiene and they include adenosine triphosphate (ATP) bioluminescence,64, 65 fluorescent markers,66, 67 and microbiologic sampling.65 Studies have demonstrated suboptimal cleaning by aerobic colony counts as well as the use of the ATP bioluminescence and fluorescent markers.64, 66 ATP bioluminescence and fluorescent markers are preferred to aerobic plate counts because they provide an immediate assessment of cleaning effectiveness.

Disinfection of Health Care Equipment and Surfaces

A great number of disinfectants are used alone or in combinations (e.g., hydrogen peroxide and peracetic acid) in the health care setting. These include alcohols, chlorine and chlorine compounds, formaldehyde, glutaraldehyde, ortho-phthalaldehyde, standard and improved hydrogen peroxide, iodophors, peracetic acid, phenolics, and quaternary ammonium compounds. With some exceptions (e.g., ethanol or bleach), commercial formulations based on these chemicals are considered unique products and must be registered with the EPA or cleared by the FDA. In most instances, a given product is designed for a specific purpose and is to be used in a certain manner. Therefore, the label should be read carefully to ensure that the right product is selected for the intended use and applied in an efficient manner. Additionally, caution must be exercised to avoid hazards with the use of cleaners and disinfectants on electronic medical equipment. Problems associated with the inappropriate use of liquids on electronic medical equipment have included equipment fires, equipment malfunctions, and health care worker burns.68

Disinfectants are not interchangeable and an overview of the performance characteristics of each is provided in the next section so the user has sufficient information to select an appropriate disinfectant for any item and use it in the most efficient way. It should be recognized that excessive costs may be attributed to incorrect concentrations and inappropriate disinfectants. Finally, occupational diseases among cleaning personnel have been associated with the use of several disinfectants, such as formaldehyde, glutaraldehyde, and chlorine, and precautions (e.g., gloves, proper ventilation) should be used to minimize exposure.69 Asthma and reactive airway disease may occur in sensitized individuals exposed to any airborne chemical, including germicides. Clinically important asthma may occur at levels below ceiling levels regulated by the Occupational and Safety Health Administration (OSHA) or recommended by the National Institute for Occupational Safety and Health. The preferred method of control is to eliminate the chemical (via engineering controls, or substitution) or relocate the worker.

Chemical Disinfectants

Alcohol

In the health care setting, “alcohol” refers to two water-soluble chemical compounds, the germicidal characteristics of which are generally underrated: ethyl alcohol and isopropyl alcohol.70 These alcohols are rapidly bactericidal rather than bacteriostatic against vegetative forms of bacteria; they also are tuberculocidal, fungicidal, and virucidal but do not destroy bacterial spores. Their cidal activity drops sharply when diluted below 50% concentration, and the optimal bactericidal concentration is in the range of 60% to 90% solutions in water (volume/volume).71, 72

Alcohols are not recommended for sterilizing medical and surgical materials, principally because of their lack of sporicidal action and their inability to penetrate protein-rich materials. Fatal postoperative wound infections with Clostridium have occurred when alcohols were used to sterilize surgical instruments contaminated with bacterial spores.73 Alcohols have been used effectively to disinfect oral and rectal thermometers, computers,60 hospital pagers, scissors, cardiopulmonary resuscitation (CPR) manikins, applanation tonometers,74 external surfaces of equipment (e.g., ventilators), and stethoscopes.75 Alcohol towelettes have been used for years to disinfect small surfaces such as rubber stoppers of multiple-dose medication vials or vaccine bottles.

Alcohols are flammable and consequently must be stored in a cool, well-ventilated area. They also evaporate rapidly, and this makes extended exposure time difficult to achieve unless the items are immersed.

Chlorine and Chlorine Compounds

Hypochlorites are the most widely used of the chlorine disinfectants and are available in liquid (e.g., sodium hypochlorite) or solid (e.g., calcium hypochlorite) forms. The most prevalent chlorine products in the United States are aqueous solutions of 5.25% to 6.15% sodium hypochlorite, which usually are called household bleach. They have a broad spectrum of antimicrobial activity (i.e., bactericidal, virucidal, fungicidal, mycobactericidal, sporicidal), do not leave toxic residues, are unaffected by water hardness, are inexpensive and fast acting,74, 76 remove dried or fixed organisms and biofilms from surfaces,77 and have a low incidence of serious toxicity.78, 79 Sodium hypochlorite at the concentration used in domestic bleach (5.25% to 6.15%) may produce ocular irritation or oropharyngeal, esophageal, and gastric burns.69, 80, 81 Other disadvantages of hypochlorites include corrosiveness to metals in high concentrations (>500 ppm), inactivation by organic matter, discoloring or “bleaching” of fabrics, release of toxic chlorine gas when mixed with ammonia or acid (e.g., household cleaning agents),82 and relative stability.83

Reports have examined the microbicidal activity of a new disinfectant, “superoxidized water.” The concept of electrolyzing saline to create a disinfectant or antiseptic is appealing because the basic materials of saline and electricity are inexpensive and the end product (i.e., water) is not damaging to the environment. The main products of this “water” are hypochlorous acid (e.g., at a concentration of about 144 mg/L) and chlorine. This is also known as electrolyzed water; and, as with any germicide, the antimicrobial activity of superoxidized water is strongly affected by the concentration of the active ingredient (available free chlorine).84 The free available chlorine concentrations of different superoxidized solutions reported in the literature range from 7 to 180 ppm.84 Data have shown that freshly generated superoxidized water is rapidly effective (<2 minutes) in achieving a 5-log10 reduction of pathogenic microorganisms (i.e., M. tuberculosis, Mycobacterium chelonae, poliovirus, human immunodeficiency virus (HIV), MRSA, E. coli, Candida albicans, Enterococcus faecalis, Pseudomonas aeruginosa) in the absence of organic loading. However, the biocidal activity of this disinfectant was substantially reduced in the presence of organic material (5% horse serum).85, 86

Hypochlorites are widely used in health care facilities in a variety of settings.76 Inorganic chlorine solution is used to disinfect tonometer heads87 and for disinfection of noncritical surfaces and equipment. A 1 : 10 to 1 : 100 dilution of 5.25% to 6.15% sodium hypochlorite (i.e., household bleach)88, 89, 90, 91 or an EPA-registered tuberculocidal disinfectant18 has been recommended for decontaminating blood spills. For small spills of blood (i.e., drops of blood) on noncritical surfaces, the area can be disinfected with a 1 : 100 dilution of 5.25% to 6.15% sodium hypochlorite or an EPA-registered tuberculocidal disinfectant. Because hypochlorites and other germicides are substantially inactivated in the presence of blood,54, 92 large spills of blood require that the surface be cleaned before an EPA-registered disinfectant or a 1 : 10 (final concentration) solution of household bleach is applied. If there is a possibility of a sharps injury, there should be an initial decontamination,69, 93 followed by cleaning and terminal disinfection (1 : 10 final concentration).54 Extreme care should always be used to prevent percutaneous injury. At least 500 ppm available chlorine for 10 minutes is recommended for decontamination of CPR training manikins. Other uses in health care include as an irrigating agent in endodontic treatment and to disinfect laundry, dental appliances, hydrotherapy tanks,40 regulated medical waste before disposal,76 applanation tonometers,74 and the water distribution system in hemodialysis centers and hemodialysis machines.9, 75 Disinfection with a 1 : 10 dilution of concentrated sodium hypochlorite (i.e., bleach) has been shown to be effective in reducing environmental contamination in patient rooms and in reducing C. difficile infection rates in hospital units where there is a high endemic C. difficile infection rates or in an outbreak setting.9, 94, 95, 96, 97, 98 At our institution, we use a sporicidal solution (5000 ppm chlorine) in all C. difficile–infected patient rooms for routine daily and terminal cleaning. This is done by one application of the sporicide covering all hand contact surfaces to allow sufficient wetness for a greater than 1-minute contact time.

Chlorine has long been favored as the preferred disinfectant in water treatment. Hyperchlorination of a Legionella-contaminated hospital water system40 resulted in a dramatic decrease (30% to 1.5%) in the isolation of Legionella pneumophila from water outlets and a cessation of health care–associated legionnaires' disease in the affected unit.99, 100 Chloramine T and hypochlorites have been used in disinfecting hydrotherapy equipment.75

Hypochlorite solutions in tap water at a pH greater than 8 stored at room temperature (23° C) in closed, opaque plastic containers may lose up to 40% to 50% of their free available chlorine level over a period of 1 month. Thus, if a user wished to have a solution containing 500 ppm of available chlorine at day 30, a solution containing 1000 ppm of chlorine should be prepared at time 0. There is no decomposition of sodium hypochlorite solution after 30 days when stored in a closed brown bottle.83

Glutaraldehyde

Glutaraldehyde is a saturated dialdehyde that has gained wide acceptance as a high-level disinfectant and chemical sterilant.101 Aqueous solutions of glutaraldehyde are acidic and generally in this state are not sporicidal. Only when the solution is “activated” (made alkaline) by use of alkalinizing agents to pH 7.5 to 8.5 does the solution become sporicidal. Once “activated” these solutions have a shelf life of minimally 14 days because of the polymerization of the glutaraldehyde molecules at alkaline pH levels. This polymerization blocks the active sites (aldehyde groups) of the glutaraldehyde molecules that are responsible for its biocidal activity.

Novel glutaraldehyde formulations (e.g., glutaraldehyde-phenol-sodium phenate, potentiated acid glutaraldehyde, stabilized alkaline glutaraldehyde) produced in the past 40 years have overcome the problem of rapid loss of activity (e.g., now use life of 28 to 30 days) while generally maintaining excellent microbicidal activity.74, 75, 102, 103 However, it should be recognized that antimicrobial activity is dependent not only on age but also on use conditions such as dilution and organic stress. The use of glutaraldehyde-based solutions in health care facilities is common because of their advantages, which include excellent biocidal properties; activity in the presence of organic matter (20% bovine serum); and noncorrosive action to endoscopic equipment, thermometers, rubber, or plastic equipment. The advantages, disadvantages, and characteristics of glutaraldehyde are listed in Table 301-2.

The in vitro inactivation of microorganisms by glutaraldehydes has been extensively investigated and reviewed.104 Several investigators showed that 2% or greater aqueous solutions of glutaraldehyde, buffered to pH 7.5 to 8.5 with sodium bicarbonate, were effective in killing vegetative bacteria in less than 2 minutes; M. tuberculosis, fungi, and viruses in less than 10 minutes; and spores of Bacillus and Clostridium species in 3 hours.104 Spores of C. difficile are more rapidly killed by 2% glutaraldehyde than are spores of other species of Clostridium and Bacillus, 105, 106 and this includes the hypervirulent binary toxin strains of C. difficile spores (W.A. Rutala, unpublished data, December 2012). There have been reports of microorganisms with relative resistance to glutaraldehyde, including some mycobacteria (M. chelonae, Mycobacterium avium-intracellulare, Mycobacterium xenopi),107, 108, 109 Methylobacterium mesophilicum, 110 Trichosporon, fungal ascospores (e.g., Microascus cinereus, Chaetomium globosum), and Cryptosporidium. 111 M. chelonae persisted in a 0.2% glutaraldehyde solution used to store porcine prosthetic heart valves,112 and a large outbreak of Mycobacterium massiliense infections in Brazil after videolaparoscopy equipment used for different elective cosmetic procedures (e.g., liposuction) was highly tolerant to 2% glutaraldehyde.113 Porins may have a role in the resistance of mycobacteria to glutaraldehyde and ortho-phthalaldehyde.114

Dilution of glutaraldehyde during use commonly occurs, and studies show a glutaraldehyde concentration decline after a few days of use in an automatic endoscope washer.115 This decline occurs because instruments are not thoroughly dried and water is carried in with the instrument, which increases the solution's volume and dilutes its effective concentration. This emphasizes the need to ensure that semicritical equipment is disinfected with an acceptable concentration of glutaraldehyde. Data suggest that 1.0% to 1.5% glutaraldehyde is the minimal effective concentration (MEC) for 2% or greater glutaraldehyde solutions when used as a high-level disinfectant.115, 116, 117 Chemical test strips or liquid chemical monitors are available for determining whether an effective concentration of glutaraldehyde is present despite repeated use and dilution. The frequency of testing should be based on how frequently the solutions are used (e.g., used daily, test daily; used weekly, test before use), but the strips should not be used to extend the use life beyond the expiration date. Data suggest the chemicals in the test strip deteriorate with time,118 and a manufacturer's expiration date should be placed on the bottles. The bottle of test strips should be dated when opened and used for the period of time indicated on the bottle (e.g., 120 days). The results of test strip monitoring should be documented. The glutaraldehyde test kits have been preliminarily evaluated for accuracy and range,118 but their reliability has been questioned.119 The concentration should be considered unacceptable or unsafe when the test indicates a dilution below the product's MEC (generally to 1.0% to 1.5% glutaraldehyde or lower) by the indicator not changing color.

Glutaraldehyde is used most commonly as a high-level disinfectant for medical equipment such as endoscopes,93 endocavitary probes, spirometry tubing, dialyzers, transducers, anesthesia and respiratory therapy equipment, hemodialysis proportioning and dialysate delivery systems, and reuse of laparoscopic disposable plastic trocars.75 Glutaraldehyde is noncorrosive to metal and does not damage lensed instruments, rubber, or plastics. The FDA-cleared labels for high-level disinfection with 2% or greater glutaraldehyde at 25° C range from 20 to 90 minutes depending on the product. However, multiple scientific studies and professional organizations support the efficacy of 2% or greater glutaraldehyde for 20 minutes at 20° C.9, 18, 37 Minimally, this latter recommendation should be followed. Glutaraldehyde should not be used for cleaning noncritical surfaces because it is too toxic and expensive.

Colitis believed to be due to glutaraldehyde exposure from residual disinfecting solution in the endoscope solution channels has been reported and is preventable by careful endoscope rinsing.69 One study found that residual glutaraldehyde levels were higher and more variable after manual disinfection (<0.2 to 159.5 mg/L) than after automatic disinfection (0.2 to 6.3 mg/L).120 Similarly, keratopathy and corneal damage were caused by ophthalmic instruments that were inadequately rinsed after soaking in 2% glutaraldehyde.121

Glutaraldehyde exposure should be monitored to ensure a safe work environment. In the absence of an OSHA permissible exposure limit, if the glutaraldehyde level is higher than the American Conference of Industrial Hygienists ceiling limit of 0.05 ppm, it would be prudent to take corrective action and repeat monitoring.122

Hydrogen Peroxide

The literature contains several accounts of the properties, germicidal effectiveness, and potential uses for stabilized hydrogen peroxide in the health care setting. Published reports ascribe good germicidal activity to hydrogen peroxide and attest to its bactericidal, virucidal, sporicidal, and fungicidal properties.123, 124, 125, 126, 127 Some other studies have shown limited bactericidal and virucidal activity of standard 3% hydrogen peroxide.58, 74 The advantages, disadvantages, and characteristics of hydrogen peroxide are listed in Table 301-2. As with other chemical sterilants, dilution of the hydrogen peroxide must be monitored by regularly testing the MEC (i.e., 7.5 to 6.0%). Compatibility testing by Olympus America of the 7.5% hydrogen peroxide found both cosmetic changes (e.g., discoloration of black anodized metal finishes)93 and functional changes with the tested endoscopes (Olympus, October 15, 1999, written communication).

Commercially available 3% hydrogen peroxide is a stable and effective disinfectant when used on inanimate surfaces. It has been used in concentrations from 3% to 6% for the disinfection of soft contact lenses (e.g., 3% for 2 to 3 hours),123, 128 tonometer biprisms, ventilators, fabrics,129 and endoscopes.130 Hydrogen peroxide was effective in spot-disinfecting fabrics in patients' rooms.129 Corneal damage from a hydrogen peroxide–soaked tonometer tip that was not properly rinsed has been reported.131

Improved Hydrogen Peroxide

An improved hydrogen peroxide–based technology has been introduced into health care for disinfection of noncritical environmental surfaces and patient equipment132 and high-level disinfection of semicritical equipment such as endoscopes.133, 134, 135 Improved hydrogen peroxide contains very low levels of anionic or nonionic surfactants or both in an acidic product that act with hydrogen peroxide to produce microbicidal activity. This combination of ingredients speeds the antimicrobial activity of hydrogen peroxide and cleaning efficiency.134, 135 Improved hydrogen peroxide is considered safe for humans and equipment and benign for the environment. In fact, improved hydrogen peroxide has the lowest EPA toxicity category (i.e., category IV) based on its oral, inhalation, and dermal toxicity, which means it is practically nontoxic and not an irritant.132, 134, 136 It is prepared and marketed by several companies in various concentrations (e.g., 0.5% to 7%), and different products may use different terminology for these products, such as “accelerated” or “activated.” Lower concentrations (i.e., 0.5%,1.4%) are designed for the low-level disinfection of noncritical environmental surfaces and patient care objects, whereas the higher concentrations can be used as high-level disinfectants for semicritical medical devices (e.g., endoscopes).

A recent study compared the bactericidal activity of a quaternary ammonium compound with two new improved hydrogen peroxide products. The improved hydrogen peroxide products were superior or similar to the quaternary ammonium compound tested. When the two improved hydrogen peroxide products were compared with standard 0.5%, 1.4%, and 3% hydrogen peroxide formulations, the improved hydrogen peroxide–based environmental surface disinfectants proved to be more effective (>6-log10 reduction) and fast-acting (30-60 seconds) microbicides in the presence of a soil load (to simulate the presence of body fluids) than commercially available hydrogen peroxide. Only 30- to 60-second contact time was studied because longer contact times (e.g., 10 minutes) are not achievable in clinical practice. Additionally, the improved hydrogen peroxide products have an EPA-registered contact time that is substantially less (e.g., 30 seconds, 1 minute for bacteria) than most EPA-registered low-level disinfectants.58 We have also recently shown that the 1.4% activated hydrogen peroxide is very effective in reducing microbial contamination of hospital privacy curtains. In our study, the activated hydrogen peroxide completely eliminated contamination with MRSA and VRE and resulted in a 98.5% reduction in microbes (only Bacillus spp. recoverable). Thus, at our institution, privacy curtains are being disinfected at the grab area by spraying the grab area of the curtain three times with activated hydrogen peroxide at discharge cleaning.

Iodophors

Iodine solutions or tinctures have long been used by health care pro­fessionals, primarily as antiseptics on skin or tissue. The FDA has not cleared any liquid chemical sterilant/high level disinfectants with iodophors as the main active ingredient. However, iodophors have been used both as antiseptics and disinfectants. An iodophor is a combination of iodine and a solubilizing agent or carrier; the resulting complex provides a sustained-release reservoir of iodine and releases small amounts of free iodine in aqueous solution. The best known and most widely used iodophor is povidone-iodine, a compound of polyvinylpyrrolidone with iodine. This product and other iodophors retain the germicidal efficacy of iodine but, unlike iodine, are generally nonstaining and are relatively free of toxicity and irritancy.137

There are several reports that documented intrinsic microbial contamination of antiseptic formulations of povidone-iodine and poloxamer-iodine.138, 139, 140 It was found that “free” iodine (I2) contributes to the bactericidal activity of iodophors and dilutions of iodophors demonstrate more rapid bactericidal action than does a full-strength povidone-iodine solution. Therefore, iodophors must be diluted according to the manufacturers' directions to achieve antimicrobial activity.

Published reports on the in vitro antimicrobial efficacy of iodophors demonstrate that iodophors are bactericidal, mycobactericidal, and virucidal but may require prolonged contact times to kill certain fungi and bacterial spores.15, 141, 142, 143, 144

Besides their use as an antiseptic, iodophors have been used for the disinfection of blood culture bottles and medical equipment such as hydrotherapy tanks and thermometers. Antiseptic iodophors are not suitable for use as hard-surface disinfectants because of concentration differences. Iodophors formulated as antiseptics contain less free iodine than those formulated as disinfectants.145 Iodine or iodine-based antiseptics should not be used on silicone catheters because the silicone tubing may be adversely affected.146

Ortho-phthalaldehyde

Ortho-phthalaldehyde (OPA) is a high-level disinfectant that received FDA clearance in October 1999. It contains at least 0.55% 1,2-benzenedicarboxaldehyde or OPA, and it has supplanted glutaraldehyde as the most commonly used “aldehyde” for high-level disinfection in the United States. OPA solution is a clear, pale-blue liquid with a pH of 7.5. The advantages, disadvantages, and characteristics of OPA are listed in Table 301-2.

Studies have demonstrated excellent microbicidal activity in in vitro studies,74, 75, 93, 111, 147, 148, 149, 150, 151, 152 including superior mycobactericidal activity (5-log10 reduction in 5 minutes) compared with glutaraldehyde. Walsh and colleagues also found OPA effective (>5-log10 reduction) against a wide range of microorganisms, including glutaraldehyde-resistant mycobacteria and Bacillus atrophaeus spores.150

OPA has several potential advantages compared with glutaraldehyde. It has excellent stability over a wide pH range (pH 3 to 9), is not a known irritant to the eyes and nasal passages, does not require exposure monitoring, has a barely perceptible odor, and requires no activation. OPA, like glutaraldehyde, has excellent material compatibility. A potential disadvantage of OPA is that it stains proteins gray (including unprotected skin) and thus must be handled with caution.93 However, skin staining would indicate improper handling that requires additional training and/or personal protective equipment (gloves, eye and mouth protection, fluid-resistant gowns). OPA residues remaining on inadequately water-rinsed transesophageal echocardiographic probes may leave stains on the patient's mouth. Meticulous cleaning, use of the correct OPA exposure time (e.g., 12 minutes), and copious rinsing of the probe with water should eliminate this problem. Because OPA has been associated with several episodes of anaphylaxis after cystoscopy,153 the manufacturer has modified its instructions for use of OPA and contraindicates the use of OPA as a disinfectant for reprocessing all urologic instrumentation for patients with a history of bladder cancer. Personal protective equipment should be worn when handling contaminated instruments, equipment, and chemicals.148 In addition, equipment must be thoroughly rinsed to prevent discoloration of a patient's skin or mucous membrane. The MEC of OPA is 0.3%, and that concentration is monitored by test strips designed specifically for the OPA solution. OPA exposure level monitoring found that the concentration during the disinfection process was significantly higher in the manual group (median, 1.43 ppb) than in the automatic group (median, 0.35 ppb). These findings corroborate other findings that show it is desirable to introduce automatic endoscope reprocessors to decrease disinfectant exposure levels among scope reprocessing technicians.154

Peracetic Acid

Peracetic, or peroxyacetic acid, is characterized by a very rapid action against all microorganisms. A special advantage of peracetic acid is its lack of harmful decomposition products (i.e., acetic acid, water, oxygen, hydrogen peroxide); it enhances removal of organic material155 and leaves no residue. It remains effective in the presence of organic matter and is sporicidal even at low temperatures. Peracetic acid can corrode copper, brass, bronze, plain steel, and galvanized iron, but these effects can be reduced by additives and pH modifications. The advantages, disadvantages, and characteristics of peracetic acid are listed in Table 301-2.

Peracetic acid will inactivate gram-positive and gram-negative bacteria, fungi, and yeasts in less than 5 minutes at less than 100 ppm. In the presence of organic matter, 200 to 500 ppm is required. For viruses the dosage range is wide (12 to 2250 ppm), with poliovirus inactivated in yeast extract in 15 minutes with 1500 to 2250 ppm. A processing system using peracetic acid at a temperature of 50° C to 56° C can be used for processing heat-sensitive semicritical and critical devices that are compatible with the peracetic acid and processing system and cannot be sterilized by other legally marketed traditional sterilization methods validated for that type of device (e.g., steam, hydrogen peroxide gas plasma, vaporized hydrogen peroxide). After processing, the devices should be used immediately or stored in a manner similar to that of a high-level disinfected endoscope.156, 157, 158 The sterilant, 35% peracetic acid, is diluted to 0.2% with tap water that has been filtered and exposed to ultraviolet light. Simulated-use trials with the earlier version of this processing system have demonstrated excellent microbicidal activity,74, 158, 159, 160, 161, 162 and three clinical trials have demonstrated both excellent microbial killing and no clinical failures leading to infection.163, 164, 165 Three clusters of infection using the earlier version of the peracetic acid automated endoscope reprocessor were linked to inadequately processed bronchoscopes when inappropriate channel connectors were used with the system.166, 167 These clusters highlight the importance of training, proper model-specific endoscope connector systems, and quality control procedures to ensure compliance with endoscope manufacturer's recommendations and professional organization guidelines. An alternative high-level disinfectant available in the United Kingdom contains 0.35% peracetic acid. Although this product is rapidly effective against a broad range of microorganisms,168, 169 it tarnishes the metal of endoscopes and is unstable, resulting in only a 24-hour use life.169

Peracetic Acid with Hydrogen Peroxide

Three chemical sterilants are FDA-cleared that contain peracetic acid plus hydrogen peroxide (0.08% peracetic acid plus 1.0% hydrogen peroxide, 0.23% peracetic acid plus 7.35% hydrogen peroxide, and 8.3% hydrogen peroxide plus 7.0% peracetic acid). The advantages, disadvantages, and characteristics of peracetic acid with hydrogen peroxide are listed in Table 301-2.

The bactericidal properties of peracetic acid plus hydrogen peroxide have been demonstrated.170 Manufacturer's data demonstrated that this combination of peracetic acid plus hydrogen peroxide inactivated all microorganisms with the exception of bacterial spores within 20 minutes. The 0.08% peracetic acid plus 1.0% hydrogen peroxide product was effective in inactivating a glutaraldehyde-resistant mycobacteria.171

The combination of peracetic acid and hydrogen peroxide has been used for disinfecting hemodialyzers.172 The percentage of dialysis centers using a peracetic acid with hydrogen peroxide–based disinfectant for reprocessing dialyzers increased from 5% in 1983 to 72% in 1997.173

Phenolics

Phenol has occupied a prominent place in the field of hospital disinfection since its initial use as a germicide by Lister in his pioneering work on antiseptic surgery. In the past 40 years, however, work has been concentrated on the numerous phenol derivatives or phenolics and their antimicrobial properties. Phenol derivatives originate when a functional group (e.g., alkyl, phenyl, benzyl, halogen) replaces one of the hydrogen atoms on the aromatic ring. Two phenol derivatives commonly found as constituents of hospital disinfectants are ortho-phenylphenol and ortho-benzyl-para-chlorophenol.

Published reports on the antimicrobial efficacy of commonly used phenolics showed that they were bactericidal, fungicidal, virucidal, and tuberculocidal.15, 53, 75, 141, 174, 175, 176, 177, 178

Many phenolic germicides are EPA registered as disinfectants for use on environmental surfaces (e.g., bedside tables, bedrails, laboratory surfaces) and noncritical medical devices. Phenolics are not FDA cleared as high-level disinfectants for use with semicritical items but could be used to preclean or decontaminate critical and semicritical devices before terminal sterilization or high-level disinfection.

The use of phenolics in nurseries has been questioned because of the occurrence of hyperbilirubinemia in infants placed in bassinets in which phenolic detergents were used.179 In addition, Doan and co-workers demonstrated bilirubin level increases in phenolic-exposed infants compared with nonphenolic-exposed infants when the phenolic was prepared according to the manufacturers' recommended dilution.180 If phenolics (or other disinfectants) are used to clean nursery floors, they must be diluted according to the recommendation on the product label. Phenolics (and other disinfectants) should not be used to clean infant bassinets and incubators while occupied. If phenolics are used to terminally clean infant bassinets and incubators, the surfaces should be rinsed thoroughly with water and dried before the infant bassinets and incubators are reused.18

Quaternary Ammonium Compounds

The quaternary ammonium compounds are widely used as surface disinfectants. There have been some reports of health care–associated infections associated with contaminated quaternary ammonium compounds used to disinfect patient care supplies or equipment such as cystoscopes or cardiac catheters.181, 182 As with several other disinfectants (e.g., phenolics, iodophors), gram-negative bacteria have been found to survive or grow in them.140

Results from manufacturers' data sheets and from published scientific literature indicate that the quaternaries sold as hospital disinfectants are generally fungicidal, bactericidal, and virucidal against lipophilic (enveloped) viruses; they are not sporicidal and generally not tuberculocidal or virucidal against hydrophilic (nonenveloped) viruses.† Poor mycobactericidal activities of quaternary ammonium compounds have been reported.49, 141

The quaternaries are commonly used in ordinary environmental sanitation of noncritical surfaces such as floors, furniture, and walls. EPA-registered quaternary ammonium compounds are appropriate to use when disinfecting medical equipment that comes into contact with intact skin (e.g., blood pressure cuffs).

Pasteurization

Pasteurization is not a sterilization process; its purpose is to destroy all pathogenic microorganisms with the exception of bacterial spores. The time-temperature relation for hot-water pasteurization is generally greater than 70° C (158° F) for 30 minutes. The water temperature and time should be monitored as part of a quality assurance program.186 Pasteurization of respiratory therapy187, 188 and anesthesia equipment189 is a recognized alternative to chemical disinfection.

Ultraviolet Light

Ultraviolet (UV) light has been recognized as an effective method for killing microorganisms. It has been suggested for use in health care for several purposes, including air disinfection, room decontamination (see “Room Decontamination,” later), surface disinfection, biofilm disinfection,190 and ultrasound probe disinfection.191 Contaminated ultrasound probes can potentially transmit pathogens. When the probe is only in contact with the patient's skin there is a low risk for infection and low-level disinfection is recommended; however, a higher level of disinfection is recommended when the probe contacts mucous membranes or nonintact skin. An evaluation of a new disinfection procedure for ultrasound probes using UV light demonstrated the median microbial reduction for UV light was 100%, 87.5% for antiseptic wiping, and 88% for dry wiping.191

Surface disinfection with UV light (100-280 nm) has been evaluated with three hospital-related surfaces, namely, aluminum (bed railings), stainless steel (operating tables), and scrubs (laboratory coats). Acinetobacter baumannii were inoculated on small coupons (103 or 105/coupon) and exposed to 90 J/m2. This exposure was effective in the inactivation of Acinetobacter from the metal coupon surfaces but ineffective in the decontamination of scrubs.192 A hand-held room decontamination technology that utilizes far-UV radiation (185 to 230 nm) to kill pathogens was evaluated and found that it rapidly kills C. difficile spores and other health care–associated pathogens on surfaces. However, the presence of organic matter reduces the efficacy of far-UV radiation, possibly explaining the more modest results observed on surfaces in hospital rooms that were not precleaned.193

Sterilization

Most medical and surgical devices used in health care facilities are made of materials that are heat stable and thus are sterilized by heat, primarily steam sterilization. However, since 1950, there has been an increase in medical devices and instruments made of materials (e.g., plastics) that require low-temperature sterilization. ETO has been used since the 1950s for heat- and moisture-sensitive medical devices. Within the past 15 years, a number of new, low-temperature sterilization systems (e.g., hydrogen peroxide gas plasma, vaporized hydrogen peroxide) have been developed and are being used to sterilize medical devices. This section reviews sterilization technologies used in health care and makes recommendations for their optimum performance in the processing of medical devices.9, 194

Sterilization destroys all microorganisms on the surface of an object or in a fluid to prevent disease transmission associated with the use of that item. Although the use of inadequately sterilized critical items represents a high risk for transmitting pathogens, documented transmission of pathogens associated with an inadequately sterilized critical item is exceedingly rare.195, 196, 197 This is likely due to the wide margin of safety associated with the sterilization processes used in health care facilities. The concept of what constitutes “sterile” is measured as a probability of sterility for each item to be sterilized. This probability is commonly referred to as the sterility assurance level (SAL) of the product and is defined as the probability of a single viable microorganism occurring on a product after sterilization. SAL is normally expressed as 10−n. For example, if the probability of a spore surviving were one in 1 million, the SAL would be 10−6.198, 199 Dual SALs (e.g., 10−3 SAL for blood culture tubes, drainage bags; 10−6 SAL for scalpels, implants) have been used in the United States for many years, and the choice of a 10−6 SAL was strictly arbitrary and not associated with any adverse outcomes (e.g., patient infections).198

Medical devices that have contact with sterile body tissues or fluids are considered critical items. These items should be sterile when used because any microbial contamination could result in disease transmission. Such items include surgical instruments, biopsy forceps, and implanted medical devices. If these items are heat resistant, the recommended sterilization process is steam sterilization, because it has the largest margin of safety due to its reliability, consistency, lethality, and least effect from organic/inorganic soils. However, reprocessing heat- and moisture-sensitive items requires use of a low-temperature sterilization technology (e.g., ETO, hydrogen peroxide gas plasma, vaporized hydrogen peroxide).200 A summary of the advantages and disadvantages for commonly used sterilization technologies is presented in Table 301-3 .

TABLE 301-3

Summary of Advantages and Disadvantages of Commonly Used Sterilization Technologies

STERILIZATION METHODADVANTAGESDISADVANTAGES
Steam Nontoxic to patient, staff, environment
Cycle easy to control and monitor
Rapidly microbicidal
Least affected by organic/inorganic soils among sterilization processes listed
Rapid cycle time
Penetrates medical packing, device lumens
Deleterious for heat-sensitive instruments
Microsurgical instruments damaged by repeated exposure
May leave instruments wet, causing them to rust
Potential for burns
Hydrogen peroxide gas plasma Safe for the environment
Leaves no toxic residuals
Cycle time is ≥24 min and no aeration necessary
Used for heat- and moisture-sensitive items since process temperature <50° C
Simple to operate, install (208-V outlet), and monitor
Compatible with most medical devices
Only requires electrical outlet
Cellulose (paper), linens, and liquids cannot be processed.
Endoscope or medical device restrictions based on lumen internal diameter and length (see manufacturer's recommendations)
Requires synthetic packaging (polypropylene wraps, polyolefin pouches) and special container tray
Hydrogen peroxide may be toxic at levels greater than 1 ppm TWA.
100% Ethylene oxide (ETO) Penetrates packaging materials, device lumens
Single-dose cartridge and negative-pressure chamber minimizes the potential for gas leak and ETO exposure
Simple to operate and monitor
Compatible with most medical materials
Requires aeration time to remove ETO residue
ETO is toxic, a carcinogen, and flammable.
ETO emission regulated by states but catalytic cell removes 99.9% of ETO and converts it to CO2 and H2O.
ETO cartridges should be stored in flammable liquid storage cabinet.
Lengthy cycle/aeration time
ETO mixtures:
8.6% ETO/91.4% HCFC
10% ETO/90% HCFC
8.5% ETO/91.5% CO2
Penetrates medical packaging and many plastics
Compatible with most medical materials
Cycle easy to control and monitor
Some states (e.g., CA, NY, MI) require ETO emission reduction of 90%-99.9%.
CFC (inert gas that eliminates explosion hazard) banned in 1995
Potential hazards to staff and patients
Lengthy cycle/aeration time
ETO is toxic, a carcinogen, and flammable.
ETO mixtures to be phased out by end of 2014
Vaporized hydrogen peroxide Safe for the environment and health care worker
Leaves no toxic residue; no aeration necessary
Fast cycle time: 55 min
Used for heat and moisture sensitive items (metal and nonmetal devices)
Medical devices restrictions based on lumen internal diameter and length; see manufacturer's recommendations (e.g., stainless steel lumen 1-mm diameter, 125-mm length)
Not used for liquid, linens, powders, or any cellulose materials
Requires synthetic packaging (polypropylene)
Limited materials compatibility data
Limited clinical use and comparative microbicidal efficacy data

Modified from references 13, 200, 278.

Steam Sterilization

Of all the methods available for sterilization, moist heat in the form of saturated steam under pressure is the most widely used and the most dependable. Steam sterilization is nontoxic, inexpensive,201 rapidly microbicidal, and sporicidal and rapidly heats and penetrates fabrics (see Table 301-3).202 Like all sterilization processes, steam sterilization has some deleterious effects on some materials, including corrosion and combustion of lubricants associated with dental handpieces,203 reduction in ability to transmit light associated with laryngoscopes,204 and increased hardening time (fivefold to sixfold) with plaster cast.205

The basic principle of steam sterilization, as accomplished in an autoclave, is to expose each item to direct steam contact at the required temperature and pressure for the specified time. Thus, there are four parameters of steam sterilization: steam, pressure, temperature, and time. The ideal steam for sterilization is dry saturated steam and entrained water (dryness fraction ≥97%).194 Pressure serves as a means to obtain the high temperatures necessary to quickly kill microorganisms. Specific temperatures must be obtained to ensure the microbicidal activity. The two common steam sterilizing temperatures are 121° C (250° F) and 132° C (270° F). These temperatures (and other high temperatures) must be maintained for a minimal time to kill microorganisms. Recognized minimum exposure periods for sterilization of wrapped health care supplies are 30 minutes at 121° C in a gravity displacement sterilizer or 4 minutes at 132° C in a prevacuum sterilizer. At constant temperatures, sterilization times vary depending on the type of item (e.g., metal versus rubber, plastic, items with lumens), whether the item is wrapped or unwrapped, and the sterilizer type.

The two basic types of steam sterilizers (autoclaves) are the gravity displacement autoclave and the high-speed prevacuum sterilizer. In the former, steam is admitted at the top or the sides of the sterilizing chamber and, because the steam is lighter than air, forces air out the bottom of the chamber through the drain vent. The gravity displacement autoclaves are primarily used to process laboratory media, water, pharmaceutical products, regulated medical waste, and nonporous articles whose surfaces have direct steam contact. With gravity displacement sterilizers the penetration time into porous items is prolonged because of incomplete air elimination. The high-speed prevacuum sterilizers are similar to the gravity displacement sterilizers except they are fitted with a vacuum pump (or ejector) to ensure air removal from the sterilizing chamber and load before the steam is admitted. The advantage of a vacuum pump is that there is nearly instantaneous steam penetration even into porous loads.

Like other sterilization systems, the steam cycle is monitored by physical, chemical, and biological monitors. Steam sterilizers usually are monitored using a printout (or graphically) by measuring temperature, the time at the temperature, and pressure. Typically, chemical indicators are affixed to the outside and incorporated into the pack to monitor the temperature or time and temperature. The effectiveness of steam sterilization is monitored with a biological indicator containing spores of Geobacillus stearothermophilus (formerly Bacillus stearothermophilus). Positive spore test results are a relatively rare event and can be attributed to operator error, inadequate steam delivery,206 or equipment malfunction.

Portable steam sterilizers are used in outpatient, dental, and rural clinics. These sterilizers are designed for small instruments, such as hypodermic syringes and needles and dental instruments. The ability of the sterilizer to reach physical parameters necessary to achieve sterilization should be monitored by physical, chemical, and biologic indicators.

Steam sterilization should be used whenever possible on all critical and semicritical items that are heat and moisture resistant (e.g., steam sterilizable respiratory therapy and anesthesia equipment), even when not essential to prevent pathogen transmission. Steam sterilizers also are used in health care facilities to decontaminate microbiologic waste and sharps containers,207 but additional exposure time is required in the gravity displacement sterilizer for these items.

Immediate-Use Steam Sterilization

“Flash” steam sterilization was originally defined by Underwood and Perkins as sterilization of an unwrapped object at 132° C for 3 minutes at 27 to 28 pounds of pressure in a gravity displacement sterilizer.208 It was intended for instruments (e.g., dropped instruments) when there is insufficient time to sterilize an item by the preferred package method. The term “flash” arose out of the abbreviated time of exposure of the unwrapped instrument. Flash sterilization is an antiquated term that does not fully describe the various steam sterilization cycles now used to process items not intended to be stored for later use. Immediate use is defined as the shortest possible time between a sterilized item's removal from the sterilizer and its aseptic transfer to the sterile field. This implies that the sterilized item is used during the procedure for which it was sterilized and in a manner that minimizes its exposure to air and other environmental contaminants. The same critical reprocessing steps (e.g., cleaning, decontamination, rinsing, and aseptic transfer from the sterilizer to the point of use) must be followed. Immediate-use steam sterilization should not be used for convenience, as an alternative to purchasing sufficient instrument sets, or as a time saver.209, 210

Ethylene Oxide “Gas” Sterilization

ETO is a colorless gas that is flammable and explosive. The four essential parameters (operational ranges) are gas concentration (450 to 1200 mg/L); temperature (37° C to 63° C); relative humidity (40% to 80%; water molecules carry ETO to reactive sites); and exposure time (1 to 6 hours). These parameters influence the effectiveness of ETO sterilization.211, 212, 213, 214 Within certain limitations, an increase in gas concentration and temperature may shorten the time necessary for achieving sterilization.

The main disadvantages associated with ETO are the lengthy cycle time and its potential hazards to patients and staff; the main advantages are that it is highly penetrating and can sterilize occluded locations in medical items and can sterilize heat- or moisture-sensitive medical equipment without deleterious effects on the material used in the medical devices (see Table 301-3).212 Acute exposure to ETO may result in irritation (e.g., to skin, eyes, or gastrointestinal or respiratory tracts) and central nervous system depression.69 Chronic inhalation has been linked to the formation of cataracts, cognitive impairment, neurologic dysfunction, and disabling polyneuropathies.69 Occupational exposure in health care facilities has been linked to hematologic changes and an increased risk for spontaneous abortions and various cancers.69 ETO should be considered a known human carcinogen.215

The use of ETO evolved when few alternatives existed for sterilizing heat- and moisture-sensitive medical devices; however, favorable properties (see Table 301-3) account for its continued widespread use.216 Two ETO gas mixtures are available to replace ETO-chlorofluorocarbon (CFC) mixtures for large capacity, tank-supplied sterilizers. The ETO-carbon dioxide (CO2) mixture consists of 8.5% ETO and 91.5% CO2. This mixture has limited use in U.S. health care facilities but is sometimes used in hospitals in India and China. It is less expensive than ETO-hydrochlorofluorocarbons (HCFC), but a disadvantage is the need for pressure vessels rated for steam sterilization, because higher pressures (28-psi gauge) are required. The other mixture, which is a drop-in CFC replacement, is ETO mixed with HCFC. HCFCs are approximately 50-fold less damaging to the earth's ozone layer than are CFCs. The EPA will begin regulation of HCFC in the year 2015 and will terminate production in the year 2030. The ETO-HCFC mixtures have been provided by companies as a drop-in replacement for CFC-12 (one mixture consists of 8.6% ETO and 91.4% HCFC, and the other mixture is composed of 10% ETO and 90% HCFC) but will be phased out by the end of 2013.216 An alternative to the pressurized mixed-gas ETO systems is 100% ETO. Partly because of the events just described, the 100% ETO sterilizers that use unit-dose cartridges will become the systems for ETO use in U.S. health care facilities.

The excellent microbicidal activity of ETO has been demonstrated in several studies25, 161, 162, 217, 218, 219 and summarized in published reports.220 ETO inactivates all microorganisms, although bacterial spores (especially B. atrophaeus) are more resistant than other microorganisms. For this reason, B. atrophaeus is the recommended biologic indicator organism.

Like all sterilization processes, the effectiveness of ETO sterilization can be altered by lumen length, lumen diameter, inorganic salts, and organic materials.‡ For example, although ETO is not used commonly for reprocessing endoscopes,39 several studies have shown failure of ETO in inactivating contaminating spores in endoscope channels221 or lumen test units.25, 161, 218 Residual ETO levels averaging 66.2 ppm have been found even after the standard degassing time.130 Failure of ETO also has been observed when dental handpieces were contaminated with Streptococcus mutans and exposed to ETO.222 It is recommended that dental handpieces be steam sterilized.

ETO is used in health care facilities to sterilize critical items (and sometimes semicritical items) that are moisture or heat sensitive and cannot be sterilized by steam sterilization.

Hydrogen Peroxide Gas Plasma

New sterilization technology based on hydrogen peroxide and plasma was patented in 1987 and marketed in the United States in 1993. Gas plasmas have been referred to as the fourth state of matter (i.e., liquids, solids, gases, and gas plasmas). Gas plasmas are generated in an enclosed chamber under deep vacuum using radiofrequency or microwave energy to excite the gas (i.e., hydrogen peroxide) molecules and produce charged particles, many of which are in the form of free radicals (e.g., hydroxyl and hydroperoxyl). This system works by diffusing hydrogen peroxide into the chamber and then “exciting” the hydrogen peroxide into a plasma state. The combined use of hydrogen peroxide vapor and plasma safely and rapidly sterilizes instruments without leaving toxic residues. The biologic indicator used with this system is Geobacillus stearothermophilus spores.

This process has the ability to inactivate a broad range of microorganisms, including resistant bacterial spores. Studies have been conducted against vegetative bacteria (including mycobacteria), yeasts, fungi, viruses, and bacterial spores.25, 161, 219, 223, 224, 225, 226, 227, 228, 229 Like all sterilization processes, the effectiveness can be altered by lumen length, lumen diameter, inorganic salts, and organic materials.§

Materials and devices that cannot tolerate high temperatures and humidity, such as some plastics, electrical devices, and corrosion-susceptible metal alloys, can be sterilized by hydrogen peroxide gas plasma. This method has been compatible with most (>95%) medical devices and materials tested.230, 231

Vaporized Hydrogen Peroxide

A new low temperature sterilization system uses vaporized hydrogen peroxide to sterilize reusable metal and nonmetal devices used in health care facilities. The system is compatible with a wide range of medical instruments and materials (e.g., polypropylene, brass, polyethylene). There are no toxic by-products because only water vapor and oxygen are produced. The system is not intended to process liquids, linens, powders, or any cellulose materials. The system can sterilize instruments with diffusion-restricted spaces (e.g., scissors) and medical devices with a single stainless steel lumen based on lumen internal diameter and length (e.g., an inside diameter of 1 mm or larger and a length of 125 mm or shorter; see manufacturer's recommendations). Thus, gastrointestinal endoscopes and bronchoscopes cannot be sterilized in this system at the current time. Although this system has not been comparatively evaluated with other sterilization processes, vaporized hydrogen peroxide has been shown to be effective in killing spores, viruses, mycobacteria, fungi, and bacteria. Table 301-3 lists the advantages and disadvantages of this and other processes.

Disinfection

Reprocessing of Endoscopes

Physicians use endoscopes to diagnose and treat numerous medical disorders. Although endoscopes represent a valuable diagnostic and therapeutic tool in modern medicine and the incidence of infection associated with use has been reported as very low (about 1 in 1.8 million procedures),232 more health care–associated outbreaks have been linked to contaminated endoscopes than to any other medical device.4, 5, 6, 233 To prevent the spread of health care–associated infections, all heat-sensitive endoscopes (e.g., gastrointestinal endoscopes, bronchoscopes, nasopharyngoscopes) must be properly cleaned and at a minimum subjected to high-level disinfection after each use. High-level disinfection can be expected to destroy all microorganisms; although when high numbers of bacterial spores are present, a few spores may survive.

Recommendations for the cleaning and disinfection of endoscopic equipment have been published and should be strictly followed.9, 37, 234, 235 Unfortunately, audits have shown that personnel do not adhere to guidelines on reprocessing,236, 237, 238, 239 and outbreaks of infection continue to occur.240, 241 To ensure that reprocessing personnel are properly trained, there should be initial and annual competency testing for each individual who is involved in reprocessing endoscopic instruments.9, 38, 167, 234

In general, endoscope disinfection or sterilization with a liquid chemical sterilant or high-level disinfectant involves five steps after leak testing:

  • 1.

    Clean—mechanically clean internal and external surfaces, including brushing internal channels and flushing each internal channel with water and a enzymatic cleaner.

  • 2.

    Disinfect—immerse endoscope in high-level disinfectant (or chemical sterilant) and perfuse (eliminates air pockets and ensures contact of the germicide with the internal channels) disinfectant into all accessible channels such as the suction/biopsy channel and air/water channel and expose for a time recommended for specific products.

  • 3.

    Rinse—rinse the endoscope and all channels with sterile water, filtered water (commonly used with automated endoscope reprocessors), or tap water.

  • 4.

    Dry—rinse the insertion tube and inner channels with alcohol and dry with forced air after disinfection and before storage.

  • 5.

    Store—store the endoscope in a way that prevents recontamination and promotes drying (e.g., hung vertically).

Unfortunately, there is poor compliance with the recommendations for reprocessing endoscopes, which may result in patient exposure to bloodborne pathogens.242 In addition, there are rare instances in which the scientific literature and recommendations from professional organizations regarding the use of disinfectants and sterilants may differ from the manufacturer's label claim. One example is the contact time used to achieve high-level disinfection with 2% glutaraldehyde. Based on FDA requirements (FDA regulates liquid sterilants and high-level disinfectants used on critical and semicritical medical devices), manufacturers test the efficacy of their germicide formulations under worst-case conditions (i.e., minimum recommended concentration of the active ingredient) and in the presence of organic soil (typically 5% serum). The soil is used to represent the organic loading to which the device is exposed during actual use and that would remain on the device in the absence of cleaning. These stringent test conditions are designed to provide a margin of safety by ensuring that the contact conditions for the germicide provide complete elimination of the test bacteria (e.g., 105 to 106 M. tuberculosis in organic soil and dried on a scope) if inoculated into the most difficult areas for the disinfectant to penetrate and in the absence of cleaning. However, the scientific data demonstrate that M. tuberculosis levels can be reduced by at least 8 log10 with cleaning (4 log10) followed by chemical disinfection for 20 minutes at 20° C (4 to 6 log10).9, 37, 243 Because of these data, professional organizations (at least 14 professional organizations worldwide) that have endorsed an endoscope reprocessing guideline recommend contact conditions of 20 minutes at 20° C (or <20 minutes outside the United States) with 2% glutaraldehyde to achieve high-level disinfection that differs from that of the manufacturer's label.37, 244, 245, 246 It is important to emphasize that the FDA tests do not include cleaning, a critical component of the disinfection process. Therefore, when cleaning has been included in the test methodology, 2% glutaraldehyde for 20 minutes has been demonstrated to be effective in eliminating all vegetative bacteria.

OSHA Bloodborne Pathogen Standard

In December 1991 OSHA promulgated a standard entitled “Occupational Exposure to Bloodborne Pathogens” to eliminate or minimize occupational exposure to bloodborne pathogens.247 One component of this requirement is that all equipment and environmental and working surfaces be cleaned and decontaminated with an appropriate disinfectant after contact with blood or other potentially infectious materials. Although the OSHA standard does not specify the type of disinfectant or procedure, the OSHA original compliance document248 suggested that a germicide must be tuberculocidal to kill hepatitis B virus (e.g., phenolic, chlorine). However, in February 1997, OSHA amended its policy and stated that EPA-registered disinfectants that are labeled as effective against HIV and hepatitis B virus would be considered as appropriate disinfectants “provided such surfaces have not become contaminated with agent(s) or volumes of or concentrations of agent(s) for which higher level disinfection is recommended.” When bloodborne pathogens other than hepatitis B virus or HIV are of concern, OSHA continues to require the use of EPA-registered tuberculocidal disinfectants or hypochlorite solution (diluted 1 : 10 or 1 : 100 with water).89, 249 Recent studies demonstrate that, in the presence of large blood spills, a 1 : 10 final dilution of EPA-registered hypochlorite solution initially should be used to inactivate bloodborne viruses54, 250 to minimize risk for disease to the health care worker from percutaneous injury during the clean-up process.

Emerging Pathogens, Antibiotic-Resistant Bacteria, and Bioterrorism Agents

Emerging pathogens are of growing concern to the general public and infection control professionals. Relevant pathogens include Cryptosporidium parvum, C. difficile, severe acute respiratory syndrome (SARS)-coronavirus, Helicobacter pylori, E. coli O157:H7, HIV, hepatitis C virus (HCV), rotavirus, multidrug-resistant M. tuberculosis, human papillomavirus, norovirus, and nontuberculous mycobacteria (e.g., M. chelonae). Similarly, publications have highlighted the concern about the potential for bioterrorism.251 The Centers for Disease Control and Prevention (CDC) has categorized several agents as “high priority” because they can be easily disseminated or transmitted person to person, can cause high mortality, and are likely to cause public panic and social disruption.252 These agents include Bacillus anthracis (anthrax), Yersinia pestis (plague), variola major (smallpox), Francisella tularensis (tularemia), filoviruses (Ebola hemorrhagic fever, Marburg hemorrhagic fever); and arenaviruses (Lassa [Lassa fever], Junin [Argentine hemorrhagic fever]), and related viruses.252

With rare exceptions, the susceptibility of each of these pathogens to chemical disinfectants/sterilants has been studied and all of these pathogens (or surrogate microbes such as feline-calicivirus for noro­virus, vaccinia for variola,142 and B. atrophaeus [formerly Bacillus subtilis] for B. anthracis), are susceptible to currently available chemical disinfectants/sterilants.9, 253, 254 Standard sterilization and high-level disinfection procedures for patient care equipment (as recommended in this chapter) are adequate to sterilize or disinfect instruments or devices contaminated with blood or other body fluids from persons infected with bloodborne pathogens, emerging pathogens, and bioterrorism agents, with the exception of prions (see later). No changes in procedures for cleaning, disinfecting, or sterilizing need to be made.9

In addition, there are no data to show that antibiotic-resistant bacteria (methicillin-resistant Staphylococcus aureus [MRSA], vancomycin-resistant Enterococcus [VRE], multidrug-resistant M. tuberculosis) are less sensitive to the liquid chemical germicides that antibiotic-sensitive bacteria at currently used germicide contact conditions and concentrations.255, 256

Current Issues in Disinfection and Sterilization

Inactivation of Creutzfeldt-Jakob Disease Agent

Creutzfeldt-Jakob disease (CJD) is a degenerative neurologic disorder of humans with an incidence in the United States of approximately 1 case/million population/year.257, 258 CJD is believed to be caused by a proteinaceous infectious agent or prion. CJD is related to other human transmissible spongiform encephalopathies (TSEs) that include kuru (0 incidence, now eradicated), Gertsmann-Straussler-Sheinker (GSS) syndrome (1/40 million), and fatal insomnia syndrome (FFI) (<1/40 million). The agents of CJD and other TSEs exhibit an unusual resistance to conventional chemical and physical decontamination methods. Because the CJD agent is not readily inactivated by conventional disinfection and sterilization procedures and because of the invariably fatal outcome of CJD, the procedures for disinfection and sterilization of the CJD prion have been both conservative and controversial for many years.

The current recommendations consider inactivation data but also use epidemiologic studies of prion transmission, infectivity of human tissues, and the efficacy of removing proteins by cleaning.257, 259, 260 On the basis of scientific data, only critical (e.g., surgical instruments) and semicritical devices contaminated with high-risk tissue (i.e., brain, spinal cord, and eye tissue) from high-risk patients (e.g., known or suspected infection with CJD or other prion disease) require special prion reprocessing. A moist environment after contamination reduces the attachment of both protein and prion amyloid to the stainless steel surface so moist conditions should be maintained.261 After the device is clean, it should be sterilized by either autoclaving (i.e., steam sterilization) or using a combination of sodium hydroxide and autoclaving262 using one of the options below257:

  • Option 1—autoclave at 134° C for 18 minutes in a prevacuum sterilizer

  • Option 2—autoclave at 132° C for 1 hour in a gravity displacement sterilizer9, 257, 263

  • Option 3—immerse in 1N sodium hydroxide for 1 hour; remove and rinse in water, then transfer to an open pan and autoclave (121° C gravity displacement or 134°C porous or prevacuum sterilizer) for 1 hour

  • Option 4—immerse in 1N sodium hydroxide for 1 hour and heat in a gravity displacement at 121° C for 30 minutes, then clean and subject to routine sterilization.

Some data suggest the temperature should not exceed 134° C because the effectiveness of autoclaving may decline as the temperature is increased (e.g., 136° C, 138° C).264 Prion-contaminated medical devices that are impossible or difficult to clean should be discarded. To minimize environmental contamination, noncritical environmental surfaces should be covered with plastic-backed paper; and when contaminated with high-risk tissues, the paper should be properly discarded. Noncritical environmental surfaces (e.g., laboratory surfaces) contaminated with high-risk tissues should be cleaned and then spot decontaminated with a 1 : 10 dilution of hypochlorite solutions.257

Role of Surfaces in Disease Transmission

There is excellent evidence in the scientific literature that environmental contamination plays an important role in the transmission of several key health care–associated pathogens, including MRSA, VRE, Acinetobacter, norovirus, and C. difficile. 265, 266, 267, 268 All these pathogens have been demonstrated to persist in the environment for days (in some cases months), frequently contaminate the environmental surfaces in rooms of colonized or infected patients, transiently colonize the hands of health care personnel, be transmitted by health care personnel, and cause outbreaks in which environmental transmission was deemed to play a role. Importantly, a recent study by Steifel and associates demonstrated that contact with the environment was just as likely to contaminate the hands of health care workers as was direct contact with the patient.269 Further, admission to a room in which the previous patient was colonized or infected with MRSA, VRE, Acinetobacter, or C. difficile has been shown to be a risk factor for the newly admitted patient to develop colonization or infection.270, 271

Adequacy of Room Cleaning and Disinfection Using Chemical Germicides

It has long been recommended in the United States that environmental surfaces in patient rooms be cleaned and disinfected on a regular basis (e.g., daily or three times per week), when surfaces are visibly soiled, and after patient discharge (terminal cleaning).9 Disinfection is generally performed using an EPA-registered hospital disinfectant such as a quaternary ammonium compound. Recent studies have demonstrated that adequate environment cleaning is frequently lacking. For example, Carling and co-workers assessed the thoroughness of terminal cleaning in the patient's immediate environment in 23 acute care hospitals (1119 patient rooms) by using a transparent, easily cleaned, stable solution that fluoresces when exposed to hand-held UV light.272 The overall thoroughness of cleaning, expressed as a percent of surfaces evaluated, was 49% (range for all hospitals, 35% to 81%). Using a similar design, Carling and co-workers assessed the environmental cleaning in intensive care unit rooms in 16 hospitals (2320 objects) and demonstrated that only 57.1% of sites were cleaned after discharge of the room's occupant.273 A recent study using ATP bioluminescence assays and aerobic cultures demonstrated that medical equipment frequently had not been disinfected as per protocol.274

Improving Room Cleaning and Disinfection and Demonstrating the Effectiveness of Surface Decontamination in Reducing Health Care–Associated Infections

Investigators have reported that intervention programs aimed at environmental services workers resulted in significant improvement in cleaning practices.65, 66 Such interventions have generally included multiple activities: improved education, monitoring the thoroughness of cleaning (e.g., by use of ATP assays or fluorescent dyes) with feedback of performance to the environmental service workers or use of cleaning checklists or both. We have found that assignment of cleaning responsibility (e.g., medical equipment to be cleaned by nursing; environmental surfaces to be cleaned by environmental service) is also important to ensure all objects and surfaces are decontaminated, especially the surfaces of medical equipment (e.g., cardiac monitors). Improved environmental cleaning has been demonstrated to reduce the environ­mental contamination with VRE,275, 276 MRSA,276 and C. difficile. 277 Importantly, no study has reported in the postintervention period proper cleaning of more than 85% of objects. Further, all studies have only focused improvement on a limited number of “high risk” objects. Thus, a concern of published studies is that they have only demonstrated improved cleaning of a limited number of “high risk” objects (or “targeted” objects), not an improvement in the overall thoroughness of room decontamination.

“No Touch” Methods for Room Decontamination

As noted earlier, multiple studies have demonstrated that environmental surfaces and objects in rooms are frequently not properly cleaned and these surfaces may be important in transmission of health care–associated pathogens. Further, although interventions aimed at improving cleaning thoroughness have demonstrated effectiveness, many surfaces remain inadequately cleaned and therefore potentially contaminated. For this reason, several manufacturers have developed room disinfection units that can decontaminate environmental surfaces and objects. These systems use one of two methods—either ultraviolet light or hydrogen peroxide.268 These technologies supplement, but do not replace, standard cleaning and disinfection because surfaces must be physically cleaned of dirt and debris.

Ultraviolet Light for Room Decontamination

UV irradiation has been used for the control of pathogenic microorganisms in a variety of applications, such as control of legionellosis, as well as disinfection of air, surfaces, and instruments.278, 279 At certain wavelengths, UV light will break the molecular bonds in DNA, thereby destroying the organism. UV-C has a characteristic wavelength of 200 to 270 nm (e.g., 254 nm), which lies in the germicidal active portion of the electromagnetic spectrum of 200 to 320 nm. The efficacy of UV irradiation is a function of many different parameters such as intensity, exposure time, lamp placement, and air movement patterns.

An automated mobile UV-C unit has been shown to eliminate more than 3-log10 vegetative bacteria (MRSA, VRE, Acinetobacter baumannii) and more than 2.4-log10 C. difficile seeded onto Formica surfaces in patients' rooms experimentally contaminated.280 Boyce and colleagues report the results of assessing the effectiveness of the same UV-C unit to reduce environmental contamination with vegetative bacteria (measured using aerobic colony counts) and C. difficile inoculated onto stainless steel carrier disks.281 Room decontamination with the UV system resulted in significant reductions in aerobic bacteria on five high-touch surfaces. Mean C. difficile log10 reductions ranged from 1.8 to 2.9 using cycle times of 34.2 to 100.1 minutes. Surfaces in direct line-of-sight were significantly more likely to yield negative cultures after UV decontamination than before decontamination. Nerandzic and colleagues showed that UV-C at a reflected dose of 22,000 mWs/cm2 for approximately 45 minutes consistently reduced recovery of C. difficile spores and MRSA by more than 2- to 3-log10 colony-forming units (CFU)/cm2 and of VRE by more than 3- to 4-log10 CFU/cm2.282 Thus, there are now three studies that have demonstrated that a UV-C system is capable of reducing vegetative bacteria inoculated on a carrier by more than 3- to 4-log10 in 15 to 20 minutes and C. difficile by more than 1.7- to 4-log10 in 35 to 100 minutes. The studies also demonstrate reduced effectiveness when surfaces were not in direct line of sight.280, 281, 282

Hydrogen Peroxide Systems for Room Decontamination

Several systems that produce hydrogen peroxide (e.g., vapor, aerosolized dry mist) have been studied for their ability to decontaminate environmental surfaces and objects in hospital rooms. Hydrogen peroxide vapor (HPV) has been used increasingly for the decontamination of rooms in health care facilities.283, 284, 285, 286, 287, 288, 289, 290, 291, 292 Investigators found that hydrogen peroxide systems are a highly effective method for eradicating various pathogens (e.g., MRSA, M. tuberculosis, Serratia, C. difficile spores, Clostridium botulinum spores) from rooms, furniture, and equipment. Importantly, using a before-after study design, Boyce and co-workers have shown that use of hydrogen peroxide vapor was associated with a significant reduction in the incidence of C. difficile infection on five high-incidence wards.283

Comparison of Ultraviolet Irradiation versus Hydrogen Peroxide for Room Decontamination

The UV-C system studied and the systems that use hydrogen peroxide have their own advantages and disadvantages268 and there is now ample evidence that these “no-touch” systems can reduce environmental contamination with health care–associated pathogens. However, each specific system should be studied and its efficacy demonstrated before being introduced into health care facilities. The main advantage of both units is their ability to achieve substantial reductions in vegetative bacteria. As noted earlier, manual cleaning has been demonstrated to be suboptimal because many environmental surfaces are not cleaned. Another advantage is their ability to substantially reduce C. difficile because low-level disinfectants (e.g., quaternary ammonium compounds) have limited or no measurable activity against spore-forming bacteria.278 Both systems are residual free and they decontaminate all exposed surfaces and equipment in the room.

The major disadvantages of both decontamination systems are the substantial capital equipment costs, the need to remove personnel and patients from the room, thus limiting their use to terminal room disinfection (must prevent/minimize exposure to UV and hydrogen peroxide), the staff time needed to transport the system to rooms to be decontaminated and monitor its use, the need to physically clean the room of dust and debris, and the sensitivity to use parameters. There are several important differences between the two systems. The UV-C system offers faster decontamination that reduces the “down” time of the room before another patient can be admitted. The hydrogen peroxide systems have been demonstrated to be more effective in eliminating spore-forming organisms. Whether this improved sporicidal activity is clinically important is unclear because studies have demonstrated that although environmental contamination is common in the rooms of patients with C. difficile infection, the level of contamination is relatively low (also true for MRSA and VRE). Finally, the hydrogen peroxide system was demonstrated to reduce C. difficile incidence in a clinical study, whereas similar studies with the UV-C system have not been published. If additional studies continue to demonstrate a benefit, then widespread adoption of these technologies should be considered for terminal room disinfection of certain patient rooms (e.g., contact precautions) in health care facilities.

Control of Hospital Waste

Health care facilities that generate medical, chemical, or radiologic waste have a moral and legal obligation to dispose of these wastes in a manner that poses minimal potential hazard to the environment or public health. The proper disposal of these wastes requires a dynamic waste management plan that conforms to federal, state, and local regulations and provides adequate personnel and financial resources to ensure implementation.

Medical waste disposal has been as a major problem in the United States for the past 40 years. The problem has developed as a result of medical waste washing ashore in some coastal states in 1987 and 1988 and the perceived threat of acquiring HIV infection via this waste. This has led to restrictive rules governing the disposal of medical waste in many states and an increase in the volume of waste defined as regulated medical waste. Coincidentally, with an increase in volume of regulated medical waste (formerly called “infectious waste”), the options for medical waste treatment and disposal are diminishing because of space and environmental concerns. This section will review some of the principles associated with medical waste management, but a more detailed description of collection, storage, processing, transporting, treatment, and public health implications of medical waste may be found elsewhere.293, 294, 295, 296, 297

Despite the attention given to medical waste by the public, the media, and all levels of government, the terms hospital waste, medical waste, regulated medical waste, and infectious waste are often used synonymously. Hospital waste refers to all waste, biologic or nonbiologic, that is discarded and not intended for further use. Medical waste refers to materials generated as a result of patient diagnosis, immunization, or treatment, such as soiled dressings or intravenous tubing. Infectious waste refers to that portion of medical waste that could potentially transmit an infectious disease. Congress and the EPA used the term regulated medical waste rather than infectious waste in the Medical Waste Tracking Act (MWTA) of 1988 in deference to the remote possibility of disease transmission associated with this waste. Thus, medical waste is a subset of hospital waste, and regulated medical waste (which is synonymous with infectious waste from a regulatory perspective) is a subset of medical waste. 293

As stated, regulated medical waste (or infectious waste) is capable of producing an infectious disease. This definition requires a consideration of the factors necessary for disease induction that include dose, host susceptibility, presence of a pathogen, virulence of a pathogen, and the most commonly absent factor, a portal of entry. For a waste to be infectious, therefore, it must contain pathogens with sufficient virulence and quantity so that exposure to the waste by a susceptible host could result in an infectious disease. Because there are no tests that allow infectious waste to be identified objectively, responsible agencies (e.g., the CDC, EPA, or states) define waste as infectious when it is suspected to contain pathogens in sufficient number to cause disease. Not only does this subjective definition result in conflicting opinions from the CDC, EPA, and state agencies on what constitutes infectious waste and how it should be treated, but it also gives undue emphasis to the mere presence of pathogens.

Guidelines produced by the CDC have designated five types of hospital waste as regulated medical waste (i.e., microbiology laboratory waste, pathology and anatomy waste, contaminated animal carcasses, blood, and sharps).40 The EPA guidelines consider the same types of waste as infectious or regulated medical waste but also designate communicable disease isolation waste.296 In the MWTA, the EPA modified its position on “communicable disease isolation waste” by including only certain “highly” communicable disease waste such as Class 4 (e.g., Marburg, Ebola, and Lassa viruses) as regulated medical waste298 (Table 301-4 ). In a systematic random survey of all U.S. hospitals conducted in July 1987 and January 1988, the overall compliance rates with the CDC and EPA recommendations were 82% and 75%, respectively. Not only were the majority of hospitals in compliance, but the hospitals frequently treated other hospital waste as infectious, including contaminated laboratory waste (87%), surgery waste (78%), dialysis waste (69%), items contacting secretions (63%), intensive care (37%), and emergency department waste (41%).293

TABLE 301-4

Types of Medical Waste Designated as Infectious (or Regulated Medical Waste) and Recommended Disposal/Treatment Methods: CDC and EPA

SOURCE/TYPE OF MEDICAL WASTECDC
EPA
MWTA
Infectious Waste MethodsDisposal/TreatmentInfectious Waste MethodsDisposal/TreatmentInfectious Waste*
Microbiologic (e.g., stocks and cultures of infectious agents) Yes† S, I Yes S, I, TI, C Yes
Blood and blood products Yes S, I, Sew Yes S, I, Sew, C Yes
Pathologic (e.g., tissue, organs) Yes I Yes I, SW, CB Yes
Sharps (e.g., needles) Yes S, I Yes S, I Yes‡
Communicable disease isolation No Yes S, I Yes‡
Contaminated animal carcasses, body parts, and bedding Yes S, I (carcasses) Yes I, SW (not bedding) Yes
Contaminated laboratory wastes No Optional§ If considered IW, use S or I No
Surgery and autopsy wastes No Optional If considered IW, use S or I No
Dialysis Unit No Optional If considered IW, use S or I No
Contaminated equipment No Optional If considered IW, use Sor I No

Modified from Rutala WA, Mayhall CG; Society of Hospital Epidemiology of America. Position paper: Medical waste. Infect Control Hosp Epidemiol. 1992:13;38-48.

A key component in evaluating the impact of a medical waste management program is the quantity of waste produced per patient. Hospitalized patients generate about 15 pounds of hospital waste per day. The amount of hospital waste generated by U.S. hospitals is approximately 6700 tons per day. U.S. hospitals designate approximately 15% of the total hospital waste by weight as infectious (about 1000 tons of infectious waste per day).293 Not surprisingly, the percentage of medical waste treated as infectious increases with the number and types of medical waste classified as infectious. For example, about 6% of hospital waste would be treated as infectious waste if the CDC guidelines are followed but 45% of hospital waste could be considered infectious waste under the MWTA.293, 299

The vast majority of U.S. hospitals designate and treat microbiologic, pathologic, isolation, blood, and sharp waste as infectious.293 In the late 1980s, treatment of infectious waste by U.S. hospitals was most commonly accomplished by incineration (range, 64% to 93%, depending on the type of waste), but emission regulations that limit air pollutants has reduced the number and use of incineration for medical waste. For example, in September 1997 there were an estimated 2373 medical waste incinerators in the United States, but based on the EPA's 2010 inventory there are 54 infectious waste incinerators.300 Autoclaves or steam sterilizers have become the primary nonincineration technology used by hospitals to process their regulated medical waste (except pathology waste) (E. Krisiunas, written communication, 2008). Several other nonincineration alternatives have been proposed for treating regulated medical waste (e.g., mechanical/chemical disinfection, microwave decontamination, steam disinfection, and compacting).297 Nonregulated medical waste is generally discarded in a properly sited and operated sanitary landfill because this is a safe and inexpensive disposal method (e.g., landfill disposal costs $0.02 to 0.05 per pound for nonregulated medical waste versus a contract incinerator cost of $0.20 to 0.60 per pound for regulated medical waste).

The conflicting opinions of state and federal regulations are related to the paucity of microbiologic and epidemiologic evidence that medical waste represents a threat to the public health. First, with the exception of “sharps” such as needles, which have caused disease only in an occupational setting, there is no scientific evidence that medical waste has caused disease in the hospital or the community. Second, data demonstrate that household waste contains on average 100 times as many microorganisms with pathogenic potential for humans than medical waste.301 Third, detailed reports of the beach washups found that the vast majority of waste on beaches was debris (about 99%) such as wood, plastic, and paper, not medical waste. EPA documents acknowledge that much of the medical waste that washed ashore in the summer of 1988 was syringe related (65%) and came from home health care and illegal drug use. Fourth, studies have shown that most U.S. hospitals are in compliance with the CDC infectious waste guidelines. Fifth, although the principal purpose of the MWTA was to reduce medical waste on beaches, it has not demonstrated its intended benefit. The relative number of syringes on the beaches in the MWTA states was significantly greater during implementation of the Act (17.23%) than before the Act went into effect (3.2%).299 If regulatory control were based on epidemiologic, microbiologic, and environmental data, only two types of medical waste would require special handling and treatment—sharps and microbiologic waste.

Federal medical waste regulations have been promulgated by the U.S. Department of Transportation and OSHA. The Department of Transportation regulation involves the transport of infectious substances and medical waste and went into effect January 1996.302 The OSHA Bloodborne Pathogen Standard requires labeling to designate waste that poses a health threat in the workplace. The OSHA definition of regulated waste is not intended to designate waste that must be treated. In fact, generators who apply the OSHA definition of regulated waste (rather than state regulations) to designate infectious waste for treatment by incineration or other means may unintentionally incur additional expenses.247

Conclusion

When properly used, disinfection and sterilization can ensure the safe use of invasive and noninvasive medical devices. However, current disinfection and sterilization guidelines must be strictly followed.

Footnotes

*References 9, 14, 16, 18, 20, 21.

†References 15, 49, 50, 52, 53, 141, 183, 184, 185.

‡References 25, 161, 162, 218, 219, 221.

§References 25, 161, 219, 221, 223, 224.

Key References

The complete reference list is available online at Expert Consult.

10. Rutala WA, Weber DJ. Selection and use of disinfectants in healthcare. In: Mayhall CG, editor. Hospital Epidemiology and Infection Control. 4th ed. Wolters Kluwer/Lippincott Williams & Wilkins; Philadelphia: 2012. [Google Scholar]

12. Rutala WA, Weber DJ. An overview of disinfection and sterilization. In: Rutala WA, editor. Disinfection, Sterilization and Antisepsis: Principles, Practices, Current Issues, New Research, and New Technologies. Association for Professionals in Infection Control and Epidemiology; Washington, DC: 2010. pp. 18–83. [PubMed] [Google Scholar]

13. Rutala WA, Weber DJ. Disinfection and sterilization in health-care facilities. In: Jarvis WR, editor. Bennett and Brachman's Hospital Infections. 6th ed. Wolters Kluwer/Lippincott Williams & Wilkins; Philadelphia: 2013. [Google Scholar]

24. U.S. Food and Drug Administration FDA-Cleared Sterilant and High-Level Disinfectants with General Claims for Processing Reusable Medical and Dental Devices—March 2009. http://www.fda.gov/cdrh/ode/germlab.html Available at.

25. Alfa MJ, DeGagne P, Olson N. Comparison of ion plasma, vaporized hydrogen peroxide and 100% ethylene oxide sterilizers to the 12/88 ethylene oxide gas sterilizer. Infect Control Hosp Epidemiol. 1996;17:92–100. [PubMed] [Google Scholar]

26. Rutala WA, Weber DJ. New developments in reprocessing semicritical devices. Am J Infect Control. 2013;41(5 suppl):S60–S66. [PubMed] [Google Scholar]

37. Petersen BT, Chennat J, Cohen J. Multisociety guideline on reprocessing flexible GI endoscopes: 2011. Infect Control Hosp Epidemiol. 2011;32:527–537. [PubMed] [Google Scholar]

40. Centers for Disease Control and Prevention Guidelines for environmental infection control in health-care facilities, 2003. MMWR Recomm Rep. 2003;52(RR-10):1–44. [PubMed] [Google Scholar]

48. Rutala WA, Barbee SL, Aguiar NC. Antimicrobial activity of home disinfectants and natural products against potential human pathogens. Infect Control Hosp Epidemiol. 2000;21:33–38. [PubMed] [Google Scholar]

49. Best M, Sattar SA, Springthorpe VS. Efficacies of selected disinfectants against Mycobacterium tuberculosis. J Clin Microbiol. 1990;28:2234–2239. [PMC free article] [PubMed] [Google Scholar]

54. Weber DJ, Barbee SL, Sobsey MD. The effect of blood on the antiviral activity of sodium hypochlorite, a phenolic, and a quaternary ammonium compound. Infect Control Hosp Epidemiol. 1999;20:821–827. [PubMed] [Google Scholar]

58. Rutala WA, Gergen MF, Weber DJ. Efficacy of improved hydrogen peroxide against important healthcare-associated pathogens. Infect Control Hosp Epidemiol. 2012;33:1159–1161. [PubMed] [Google Scholar]

66. Carling PC, Parry MF, Rupp ME. Improving cleaning of the environment surrounding patients in 36 acute care hospitals. Infect Control Hosp Epidemiol. 2008;29:1035–1041. [PubMed] [Google Scholar]

74. Rutala WA, Peacock JE, Gergen MF. Efficacy of hospital germicides against adenovirus 8, a common cause of epidemic keratoconjunctivitis in health care facilities. Antimicrob Agents Chemother. 2006;50:1419–1424. [PMC free article] [PubMed] [Google Scholar]

76. Rutala WA, Weber DJ. Uses of inorganic hypochlorite (bleach) in health-care facilities. Clin Microbiol Rev. 1997;10:597–610. [PMC free article] [PubMed] [Google Scholar]

83. Rutala WA, Cole EC, Thomann CA. Stability and bactericidal activity of chlorine solutions. Infect Control Hosp Epidemiol. 1998;19:323–327. [PubMed] [Google Scholar]

95. Perez J, Springthorpe S, Sattar SA. Activity of selected oxidizing microbicides against spores of Clostridium difficile: relevance to environmental control. Am J Infect Control. 2005;33:320–325. [PubMed] [Google Scholar]

96. Orenstein R, Aronhalt KC, McManus JE., Jr A targeted strategy to wipe out Clostridium difficile. Infect Control Hosp Epidemiol. 2011;32:1137–1139. [PubMed] [Google Scholar]

105. Rutala WA, Gergen MF, Weber DJ. Inactivation of Clostridium difficile spores by disinfectants. Infect Control Hosp Epidemiol. 1993;14:36–39. [PubMed] [Google Scholar]

115. Mbithi JN, Springthorpe VS, Sattar SA. Bactericidal, virucidal, and mycobactericidal activities of reused alkaline glutaraldehyde in an endoscopy unit. J Clin Microbiol. 1993;31:2988–2995. [PMC free article] [PubMed] [Google Scholar]

140. Weber DJ, Rutala WA, Sickbert-Bennett EE. Outbreaks associated with contaminated antiseptics and disinfectants. Antimicrob Agents Chemother. 2007;51:916–919. [PMC free article] [PubMed] [Google Scholar]

141. Rutala WA, Cole EC, Wannamaker NS. Inactivation of Mycobacterium tuberculosis and Mycobacterium bovis by 14 hospital disinfectants. Am J Med. 1991;91:267S–271S. [PubMed] [Google Scholar]

142. Klein M, DeForest A. The inactivation of viruses by germicides. Chem Specialists Manuf Assoc Proc. 1963;49:116–118. [Google Scholar]

152. Roberts CG, Chan-Myers HB, Favero MS. Virucidal activity of ortho-phthalaldehyde solutions against hepatitis B and C viruses. Am J Infect Control. 2008;36:223–226. [PubMed] [Google Scholar]

161. Alfa MJ, Olson N, Degagne P. New low temperature sterilization technologies: microbicidal activity and clinical efficacy. In: Rutala WA, editor. Disinfection, Sterilization, and Antisepsis in Healthcare. Polyscience Publications; Champlain, NY: 1998. pp. 67–78. [Google Scholar]

174. Rutala WA, Cole EC. Ineffectiveness of hospital disinfectants against bacteria: a collaborative study. Infect Control. 1987;8:501–506. [PubMed] [Google Scholar]

198. Favero MS. Sterility assurance: concepts for patient safety. In: Rutala WA, editor. Disinfection, Sterilization and Antisepsis: Principles and Practices in Healthcare Facilities. Association for Professional in Infection Control and Epidemiology; Washington, DC: 2001. pp. 110–119. [Google Scholar]

209. Seavey R. Immediate use steam sterilization: moving beyond current policy. Am J Infect Control. 2013;41(5 suppl):S46–S48. [PubMed] [Google Scholar]

214. Association for the Advancement of Medical Instrumentation . AAMI; Arlington, VA: 2008. Ethylene Oxide Sterilization in Health Care Facilities: Safety and Effectiveness. Report No. ANSI/AAMI ST41. [Google Scholar]

242. Rutala WA, Weber DJ. How to assess risk of disease transmission when there is a failure to follow recommended disinfection and sterilization principles. Infect Control Hosp Epidemiol. 2007;28:519–524. [PubMed] [Google Scholar]

243. Rutala WA, Weber DJ. FDA labeling requirements for disinfection of endoscopes: a counterpoint. Infect Control Hosp Epidemiol. 1995;16:231–235. [PubMed] [Google Scholar]

247. Occupational Safety and Health Administration Occupational exposure to bloodborne pathogens; final rule. Fed Register. 1991;56:64003–64182. [Google Scholar]

255. Rutala WA, Stiegel MM, Sarubbi FA. Susceptibility of antibiotic-susceptible and antibiotic-resistant hospital bacteria to disinfectants. Infect Control Hosp Epidemiol. 1997;18:417–421. [PubMed] [Google Scholar]

256. Weber DJ, Rutala WA. Use of germicides in the home and health care setting: is there a relationship between germicide use and antimicrobial resistance. Infect Control Hosp Epidemiol. 2006;27:1107–1119. [PubMed] [Google Scholar]

257. Rutala WA, Weber DJ. Guideline for disinfection and sterilization of prion-contaminated medical instruments. Infect Control Hosp Epidemiol. 2010;31:107–117. [PubMed] [Google Scholar]

259. Belay ED, Schonberger LB, Brown P. Disinfection and sterilization of prion-contaminated medical instruments. Infect Control Hosp Epidemiol. 2010;31:1304–1306. [PubMed] [Google Scholar]

260. Rutala WA, Weber DJ. Disinfection and sterilization of prion-contaminated medical instruments: reply to Belay. Infect Control Hosp Epidemiol. 2010;31:1306–1307. [Google Scholar]

265. Otter JA. The role played by contaminated surfaces in the transmission of nosocomial pathogens. Infect Control Hosp Epidemiol. 2011;32:687–699. [PubMed] [Google Scholar]

266. Boyce JM. Environmental contamination makes an important contribution to hospital infection. J Hosp Infect. 2007;65:50–54. [PubMed] [Google Scholar]

267. Weber DJ, Rutala WA, Miller MB. Role of hospital surfaces in the transmission of emerging health care-associated pathogens: norovirus, Clostridium difficile, and Acinetobacter species. Am J Infect Control. 2010;38:S25–S33. [PubMed] [Google Scholar]

268. Rutala WA, Weber DJ. Are room decontamination units needed to prevent transmission of environmental pathogens? Infect Control Hosp Epidemiol. 2011;32:743–747. [PubMed] [Google Scholar]

270. Huang SS, Datta R, Platt R. Risk of acquiring antibiotic-resistant bacteria from prior room occupants. Arch Intern Med. 2006;166:1945–1951. [PubMed] [Google Scholar]

271. Shaughnessy MK, Micielli RL, DePestal DD. Evaluation of hospital room assignment and acquisition of Clostridium difficile infection. Infect Control Hosp Epidemiol. 2011;32:201–206. [PubMed] [Google Scholar]

272. Carling PC, Parry MF, Von Beheren SM, Healthcare Environmental Hygiene Study Group Identifying opportunities to enhance environmental cleaning in 23 acute care hospitals. Infect Control Hosp Epidemiol. 2008;29:1–7. [PubMed] [Google Scholar]

280. Rutala WA, Gergen MF, Weber DJ. Room decontamination by ultraviolet radiation. Infect Control Hosp Epidemiol. 2010;31:1025–1029. [PubMed] [Google Scholar]

283. Boyce JM, Havill NL, Otter JA. Impact of hydrogen peroxide vapor room decontamination on Clostridium difficile environmental contamination and transmission in a healthcare setting. Infect Control Hosp Epidemiol. 2008;29:723–729. [PubMed] [Google Scholar]

293. Rutala WA, Odette RL, Samsa GP. Management of infectious waste by United States hospitals. JAMA. 1989;262:1635–1640. [PubMed] [Google Scholar]

294. Rutala WA, Sarubbi FA. Management of infectious waste from hospitals. Infect Control. 1983;4:198–204. [PubMed] [Google Scholar]

299. Rutala WA, Weber DJ. Infectious waste: mismatch between science and policy. N Engl J Med. 1991;325:578–582. [PubMed] [Google Scholar]

301. Rutala WA, Mayhall CG, Society for Hospital Epidemiology of America SHEA position paper: Medical waste. Infect Control Hosp Epidemiol. 1992;13:38–48. [PubMed] [Google Scholar]

References

2. Weber DJ. Managing and preventing exposure events from inappropriately reprocessed endoscopes. Infect Control Hosp Epidemiol. 2012;33:657–660. [PubMed] [Google Scholar]

3. Uttley AH, Simpson RA. Audit of bronchoscope disinfection: a survey of procedures in England and Wales and incidents of mycobacterial contamination. J Hosp Infect. 1994;26:301–308. [PubMed] [Google Scholar]

4. Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993;118:117–128. [PubMed] [Google Scholar]

5. Weber DJ, Rutala WA. Lessons from outbreaks associated with bronchoscopy. Infect Control Hosp Epidemiol. 2001;22:403–408. [PubMed] [Google Scholar]

6. Weber DJ, Rutala WA, DiMarino AJ., Jr . The prevention of infection following gastrointestinal endoscopy: the importance of prophylaxis and reprocessing. In: DiMarino AJ Jr, Benjamin SB, editors. Gastrointestinal Diseases: An Endoscopic Approach. 2nd ed. Slack Inc; Thorofare, NJ: 2002. pp. 87–106. [Google Scholar]

7. Meyers H, Brown-Elliott BA, Moore D. An outbreak of Mycobacterium chelonae infection following liposuction. Clin Infect Dis. 2002;34:1500–1507. [PubMed] [Google Scholar]

8. Lowry PW, Jarvis WR, Oberle AD. Mycobacterium chelonae causing otitis media in an ear-nose-and-throat practice. N Engl J Med. 1988;319:978–982. [PubMed] [Google Scholar]

10. Rutala WA, Weber DJ. Selection and use of disinfectants in healthcare. In: Mayhall CG, editor. Hospital Epidemiology and Infection Control. 4th ed. Wolters Kluwer/Lippincott Williams & Wilkins; Philadelphia: 2012. [Google Scholar]

11. Rutala WA, Weber DJ. Disinfection, sterilization, and control of hospital waste. In: Mandell GL, Bennett JE, Dolin R, editors. Principles and Practice of Infectious Diseases. 7th ed. Churchill Livingstone Elsevier; Philadelphia: 2009. pp. 3677–3695. [Google Scholar]

12. Rutala WA, Weber DJ. An overview of disinfection and sterilization. In: Rutala WA, editor. Disinfection, Sterilization and Antisepsis: Principles, Practices, Current Issues, New Research, and New Technologies. Association for Professionals in Infection Control and Epidemiology; Washington, DC: 2010. pp. 18–83. [PubMed] [Google Scholar]

13. Rutala WA, Weber DJ. Disinfection and sterilization in health-care facilities. In: Jarvis WR, editor. Bennett and Brachman's Hospital Infections. 6th ed. Wolters Kluwer/Lippincott Williams & Wilkins; Philadelphia: 2013. [Google Scholar]

14. Favero MS, Bond WW. Chemical disinfection of medical and surgical materials. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 881–917. [Google Scholar]

15. Spaulding EH. Chemical disinfection of medical and surgical materials. In: Lawrence C, Block SS, editors. Disinfection, Sterilization, and Preservation. Lea & Febiger; Philadelphia: 1968. pp. 517–531. [Google Scholar]

16. Simmons BP. CDC guidelines for the prevention and control of nosocomial infections: guideline for hospital environmental control. Am J Infect Control. 1983;11:97–120. [PubMed] [Google Scholar]

17. Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. [Google Scholar]

18. Rutala WA, 1994, 1995, and 1996 APIC Guidelines Committee APIC guideline for selection and use of disinfectants. Association for Professionals in Infection Control and Epidemiology, Inc. Am J Infect Control. 1996;24:313–342. [PubMed] [Google Scholar]

19. Rutala WA. Disinfection, sterilization and waste disposal. In: Wenzel RP, editor. Prevention and Control of Nosocomial Infections. 3rd ed. Williams and Wilkins; Baltimore: 1997. pp. 539–593. [Google Scholar]

20. Garner JS, Favero MS. CDC guideline for handwashing and hospital environmental control, 1985. Infect Control. 1986;7:231–243. [PubMed] [Google Scholar]

21. Rutala WA. APIC guideline for selection and use of disinfectants. Am J Infect Control. 1990;18:99–117. [PubMed] [Google Scholar]

22. McDonnell G, Burke P. Disinfection: is it time to reconsider Spaulding? J Hosp Infect. 2011;78:163–170. [PubMed] [Google Scholar]

23. Miner N, Harris V, Cao TD. Aldahol high-level disinfectant. Am J Infect Control. 2010;38:205–211. [PubMed] [Google Scholar]

24. U.S. Food and Drug Administration FDA-Cleared Sterilant and High-Level Disinfectants with General Claims for Processing Reusable Medical and Dental Devices—March 2009. http://www.fda.gov/cdrh/ode/germlab.html Available at.

25. Alfa MJ, DeGagne P, Olson N. Comparison of ion plasma, vaporized hydrogen peroxide and 100% ethylene oxide sterilizers to the 12/88 ethylene oxide gas sterilizer. Infect Control Hosp Epidemiol. 1996;17:92–100. [PubMed] [Google Scholar]

26. Rutala WA, Weber DJ. New developments in reprocessing semicritical devices. Am J Infect Control. 2013;41(5 suppl):S60–S66. [PubMed] [Google Scholar]

27. Rutala WA, Gergen MF, Weber DJ. Disinfection of a probe used in ultrasound-guided prostate biopsy. Infect Control Hosp Epidemiol. 2007;28:916–919. [PubMed] [Google Scholar]

28. Rutala WA, Gergen MF, Weber DJ. Disinfection of an infrared coagulation device used to treat hemorrhoids. Am J Infect Control. 2012;40:78–79. [PubMed] [Google Scholar]

29. Wendelboe AM, Baumbach J, Blossom DB. Outbreak of cystoscopy related infections with Pseudomonas aeruginosa: New Mexico, 2007. J Urol. 2008;180:588–592. [PubMed] [Google Scholar]

30. Foliente RL, Kovacs BJ, Aprecio RM. Efficacy of high-level disinfectants for reprocessing gastrointestinal endoscopes in simulated-use testing. Gastrointest Endosc. 2001;53:456–462. [PubMed] [Google Scholar]

31. Kovacs BJ, Chen YK, Kettering JD. High-level disinfection of gastrointestinal endoscopes: are current guidelines adequate? Am J Gastroenterol. 1999;94:1546–1550. [PubMed] [Google Scholar]

32. Wallace RJ, Jr, Brown BA, Griffith DE. Nosocomial outbreaks/pseudo-outbreaks caused by nontuberculous mycobacteria. Annu Rev Microbiol. 1998;52:453–490. [PubMed] [Google Scholar]

33. Meenhorst PL, Reingold AL, Groothuis DG. Water-related nosocomial pneumonia caused by Legionella pneumophila serogroups 1 and 10. J Infect Dis. 1985;152:356–364. [PubMed] [Google Scholar]

34. Atlas RM. Legionella: from environmental habitats to disease pathology, detection and control. Environ Microbiol. 1999;1:283–293. [PubMed] [Google Scholar]

35. Rutala WA, Weber DJ. Water as a reservoir of nosocomial pathogens. Infect Control Hosp Epidemiol. 1997;18:609–616. [PubMed] [Google Scholar]

36. Weber DJ, Rutala WA. Environmental issues and nosocomial infections. In: Wenzel RP, editor. Prevention and Control of Nosocomial Infections. 3rd ed. Williams & Wilkins; Baltimore: 1997. pp. 491–514. [Google Scholar]

37. Petersen BT, Chennat J, Cohen J. Multisociety guideline on reprocessing flexible GI endoscopes: 2011. Infect Control Hosp Epidemiol. 2011;32:527–537. [PubMed] [Google Scholar]

38. Gerding DN, Peterson LR, Vennes JA. Cleaning and disinfection of fiberoptic endoscopes: evaluation of glutaraldehyde exposure time and forced-air drying. Gastroenterology. 1982;83:613–618. [PubMed] [Google Scholar]

39. Rutala WA, Clontz EP, Weber DJ. Disinfection practices for endoscopes and other semicritical items. Infect Control Hosp Epidemiol. 1991;12:282–288. [PubMed] [Google Scholar]

40. Centers for Disease Control and Prevention Guidelines for environmental infection control in health-care facilities, 2003. MMWR Recomm Rep. 2003;52(RR-10):1–44. [PubMed] [Google Scholar]

41. Huslage K, Rutala WA, Sickbert-Bennett E. A quantitative approach to defining high-touch surfaces in hospitals. Infect Control Hosp Epidemiol. 2010;31:850–853. [PubMed] [Google Scholar]

42. Sattar SA, Lloyd-Evans N, Springthorpe VS. Institutional outbreaks of rotavirus diarrhoea: potential role of fomites and environmental surfaces as vehicles for virus transmission. J Hyg. 1986;96:277–289. [PMC free article] [PubMed] [Google Scholar]

43. Weber DJ, Rutala WA. Role of environmental contamination in the transmission of vancomycin-resistant enterococci. Infect Control Hosp Epidemiol. 1997;18:306–309. [PubMed] [Google Scholar]

44. Ward RL, Bernstein DI, Knowlton DR. Prevention of surface-to-human transmission of rotaviruses by treatment with disinfectant spray. J Clin Microbiol. 1991;29:1991–1996. [PMC free article] [PubMed] [Google Scholar]

45. Sattar SA, Jacobsen H, Springthorpe VS. Chemical disinfection to interrupt transfer of rhinovirus type 14 from environmental surfaces to hands. Appl Environ Microbiol. 1993;59:1579–1585. [PMC free article] [PubMed] [Google Scholar]

46. Gwaltney JM, Jr, Hendley JO. Transmission of experimental rhinovirus infection by contaminated surfaces. Am J Epidemiol. 1982;116:828–833. [PubMed] [Google Scholar]

47. Sattar SA, Jacobsen H, Rahman H. Interruption of rotavirus spread through chemical disinfection. Infect Control Hosp Epidemiol. 1994;15:751–756. [PubMed] [Google Scholar]

48. Rutala WA, Barbee SL, Aguiar NC. Antimicrobial activity of home disinfectants and natural products against potential human pathogens. Infect Control Hosp Epidemiol. 2000;21:33–38. [PubMed] [Google Scholar]

49. Best M, Sattar SA, Springthorpe VS. Efficacies of selected disinfectants against Mycobacterium tuberculosis. J Clin Microbiol. 1990;28:2234–2239. [PMC free article] [PubMed] [Google Scholar]

50. Best M, Kennedy ME, Coates F. Efficacy of a variety of disinfectants against Listeria spp. Appl Environ Microbiol. 1990;56:377–380. [PMC free article] [PubMed] [Google Scholar]

51. Best M, Springthorpe VS, Sattar SA. Feasibility of a combined carrier test for disinfectants: studies with a mixture of five types of microorganisms. Am J Infect Control. 1994;22:152–162. [PubMed] [Google Scholar]

52. Springthorpe VS, Grenier JL, Lloyd-Evans N. Chemical disinfection of human rotaviruses: efficacy of commercially-available products in suspension tests. J Hyg. 1986;97:139–161. [PMC free article] [PubMed] [Google Scholar]

53. Sattar SA, Springthorpe VS. Survival and disinfectant inactivation of the human immunodeficiency virus: a critical review. Rev Infect Dis. 1991;13:430–447. [PubMed] [Google Scholar]

54. Weber DJ, Barbee SL, Sobsey MD. The effect of blood on the antiviral activity of sodium hypochlorite, a phenolic, and a quaternary ammonium compound. Infect Control Hosp Epidemiol. 1999;20:821–827. [PubMed] [Google Scholar]

55. Ciesek S, Friesland M, Steinmann J. How stable is the hepatitis C virus (HCV)? Environmental stability of HCV and its susceptibility to chemical biocides. J Infect Dis. 2010;201:1859–1866. [PubMed] [Google Scholar]

56. Dellanno C, Vega Q, Boesenberg D. The antiviral action of common household disinfectants and antiseptics against murine hepatitis virus, a potential surrogate for SARS coronavirus. Am J Infect Control. 2009;37:649–652. [PMC free article] [PubMed] [Google Scholar]

57. Jeong EK, Bae JE, Kim IS. Inactivation of influenza A virus H1N1 by disinfection process. Am J Infect Control. 2010;38:354–360. [PubMed] [Google Scholar]

58. Rutala WA, Gergen MF, Weber DJ. Efficacy of improved hydrogen peroxide against important healthcare-associated pathogens. Infect Control Hosp Epidemiol. 2012;33:1159–1161. [PubMed] [Google Scholar]

59. Pentella MA, Fisher T, Chandler S. Are disinfectants accurately prepared for use in hospital patient care areas? Infect Control Hosp Epidemiol. 2000;21:103. [Google Scholar]

60. Rutala WA, White MS, Gergen MF. Bacterial contamination of keyboards: efficacy and functional impact of disinfectants. Infect Control Hosp Epidemiol. 2006;27:372–377. [PubMed] [Google Scholar]

61. Rutala WA, Gergen MF, Weber DJ. Microbiologic evaluation of microfiber mops for surface disinfection. Am J Infect Control. 2007;35:569–573. [PubMed] [Google Scholar]

62. Westwood JC, Mitchell MA, Legace S. Hospital sanitation: the massive bacterial contamination of the wet mop. Appl Microbiol. 1971;21:693–697. [PMC free article] [PubMed] [Google Scholar]

63. Cooper RA, Griffith CJ, Malik RE. Monitoring the effectiveness of cleaning in four British hospitals. Am J Infect Control. 2007;35:338–341. [PubMed] [Google Scholar]

64. Boyce JM, Havill NL, Dumigan DG. Monitoring the effectiveness of hospital cleaning practices by use of an adenosine triphosphate bioluminescence assay. Infect Control Hosp Epidemiol. 2009;30:678–684. [PubMed] [Google Scholar]

65. Lewis T, Griffith C, Gallo M. A modified ATP benchmark for evaluating the cleaning of some hospital environmental surfaces. J Hosp Infect. 2008;69:156–163. [PubMed] [Google Scholar]

66. Carling PC, Parry MF, Rupp ME. Improving cleaning of the environment surrounding patients in 36 acute care hospitals. Infect Control Hosp Epidemiol. 2008;29:1035–1041. [PubMed] [Google Scholar]

67. Blue J, O’Neill C, Speziale P. Use of a fluorescent chemical as a quality indicator for a hospital cleaning program. Can J Infect Control. 2008;23:216–219. [PubMed] [Google Scholar]

68. U.S. Food and Drug Administration. Public Health Notification from FDA, CDC, EPA and OSHA: Avoiding Hazards with Using Cleaners and Disinfectants on Electronic Medical Equipment. U.S. Department of Health & Human Services, October 31, 2007.

69. Weber DJ, Rutala WA. Occupational risks associated with the use of selected disinfectants and sterilants. In: Rutala WA, editor. Disinfection, Sterilization, and Antisepsis in Healthcare. Polyscience Publications; Champlain, NY: 1998. pp. 211–226. [Google Scholar]

70. Spaulding EH. Alcohol as a surgical disinfectant. AORN J. 1964;2:67–71. [Google Scholar]

71. Morton HE. The relationship of concentration and germicidal efficiency of ethyl alcohol. Ann NY Acad Sci. 1950;53:191–196. [PubMed] [Google Scholar]

72. Ali Y, Dolan MJ, Fendler EJ. Alcohols. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 229–254. [Google Scholar]

73. Nye RN, Mallory TB. A note on the fallacy of using alcohol for the sterilization of surgical instruments. Boston Med Surg J. 1923;189:561–563. [Google Scholar]

74. Rutala WA, Peacock JE, Gergen MF. Efficacy of hospital germicides against adenovirus 8, a common cause of epidemic keratoconjunctivitis in health care facilities. Antimicrob Agents Chemother. 2006;50:1419–1424. [PMC free article] [PubMed] [Google Scholar]

75. Rutala WA, Weber DJ. Selection and use of disinfectants in healthcare. In: Mayhall CG, editor. Hospital Epidemiology and Infection Control. 3rd ed. Lippincott Williams & Wilkins; Philadelphia: 2004. pp. 1473–1522. [Google Scholar]

76. Rutala WA, Weber DJ. Uses of inorganic hypochlorite (bleach) in health-care facilities. Clin Microbiol Rev. 1997;10:597–610. [PMC free article] [PubMed] [Google Scholar]

77. Merritt K, Hitchins VM, Brown SA. Safety and cleaning of medical materials and devices. J Biomed Mater Res. 2000;53:131–136. [PubMed] [Google Scholar]

78. Jakobsson SW, Rajs J, Jonsson JA. Poisoning with sodium hypochlorite solution: report of a fatal case, supplemented with an experimental and clinico-epidemiological study. Am J Forens Med Pathol. 1991;12:320–327. [PubMed] [Google Scholar]

79. Heidemann SM, Goetting MG. Treatment of acute hypoxemic respiratory failure caused by chlorine exposure. Pediatr Emerg Care. 1991;7:87–88. [PubMed] [Google Scholar]

80. French RJ, Tabb HG, Rutledge LJ. Esophageal stenosis produced by ingestion of bleach: report of two cases. South Med J. 1970;63:1140–1144. [PubMed] [Google Scholar]

81. Ingram TA. Response of the human eye to accidental exposure to sodium hypochlorite. J Endod. 1990;16:235–238. [PubMed] [Google Scholar]

82. Mrvos R, Dean BS, Krenzelok EP. Home exposures to chlorine/chloramine gas: review of 216 cases. South Med J. 1993;86:654–657. [PubMed] [Google Scholar]

83. Rutala WA, Cole EC, Thomann CA. Stability and bactericidal activity of chlorine solutions. Infect Control Hosp Epidemiol. 1998;19:323–327. [PubMed] [Google Scholar]

84. Sampson MN, Muir AVG. Not all super-oxidized waters are the same. J Hosp Infect. 2002;52:227–228. [PubMed] [Google Scholar]

85. Selkon JB, Babb JR, Morris R. Evaluation of the antimicrobial activity of a new super-oxidized water, Sterilox®, for the disinfection of endoscopes. J Hosp Infect. 1999;41:59–70. [PubMed] [Google Scholar]

86. Shetty N, Srinivasan S, Holton J. Evaluation of microbicidal activity of a new disinfectant: Sterilox® 2500 against Clostridium difficile spores, Helicobacter pylori, vancomycin resistant Enterococcus species, Candida albicans and several Mycobacterium species. J Hosp Infect. 1999;41:101–105. [PubMed] [Google Scholar]

87. Nagington J, Sutehall GM, Whipp P. Tonometer disinfection and viruses. Br J Ophthalmol. 1983;67:674–676. [PMC free article] [PubMed] [Google Scholar]

88. Centers for Disease Control Acquired immune deficiency syndrome (AIDS): precautions for clinical and laboratory staffs. MMWR Morb Mortal Wkly Rep. 1982;31:577–580. [PubMed] [Google Scholar]

89. Centers for Disease Control Recommendations for prevention of HIV transmission in health-care settings. MMWR Morb Mortal Wkly Rep. 1987;36:S3–S18. [Google Scholar]

90. Centers for Disease Control Guidelines for prevention of transmission of human immunodeficiency virus and hepatitis B virus to health-care and public-safety workers. MMWR Morb Mortal Wkly Rep. 1989;38(S-6):1–37. [PubMed] [Google Scholar]

91. Garner JS, Simmons BP. Guideline for isolation precautions in hospitals. Infect Control. 1983;4(4 suppl):245–325. [PubMed] [Google Scholar]

92. Bloomfield SF, Miller EA. A comparison of hypochlorite and phenolic disinfectants for disinfection of clean and soiled surfaces and blood spillages. J Hosp Infect. 1989;13:231–239. [PubMed] [Google Scholar]

93. Rutala WA, Weber DJ. Disinfection of endoscopes: review of new chemical sterilants used for high-level disinfection. Infect Control Hosp Epidemiol. 1999;20:69–76. [PubMed] [Google Scholar]

94. Gerding DN, Muto CA, Owens RC., Jr Measures to control and prevent Clostridium difficile infection. Clin Infect Dis. 2008;46(suppl 1):S43–S49. [PubMed] [Google Scholar]

95. Perez J, Springthorpe S, Sattar SA. Activity of selected oxidizing microbicides against spores of Clostridium difficile: relevance to environmental control. Am J Infect Control. 2005;33:320–325. [PubMed] [Google Scholar]

96. Orenstein R, Aronhalt KC, McManus JE., Jr A targeted strategy to wipe out Clostridium difficile. Infect Control Hosp Epidemiol. 2011;32:1137–1139. [PubMed] [Google Scholar]

97. Hacek DM, Ogle AM, Fisher A. Significant impact of terminal room cleaning with bleach on reducing nosocomial Clostridium difficile. Am J Infect Control. 2010;38:350–353. [PubMed] [Google Scholar]

98. Mayfield JL, Leet T, Miller J. Environmental control to reduce transmission of Clostridium difficile. Clin Infect Dis. 2000;31:995–1000. [PubMed] [Google Scholar]

99. Helms CM, Massanari RM, Zeitler R. Legionnaires’ disease associated with a hospital water system: a cluster of 24 nosocomial cases. Ann Intern Med. 1983;99:172–178. [PubMed] [Google Scholar]

100. Helms C, Massanari R, Wenzel R. Legionnaires’ disease associated with a hospital water system. A five-year progress report on continuous hyperchlorination. JAMA. 1988;259:2423–2427. [PubMed] [Google Scholar]

101. Cheung RJ, Ortiz D, DiMarino AJ., Jr GI endoscopic reprocessing practices in the United States. Gastrointest Endosc. 1999;50:362–368. [PubMed] [Google Scholar]

102. Miner NA, McDowell JW, Willcockson GW. Antimicrobial and other properties of a new stabilized alkaline glutaraldehyde disinfectant/sterilizer. Am J Hosp Pharm. 1977;34:376–382. [PubMed] [Google Scholar]

103. Pepper RE. Comparison of the activities and stabilities of alkaline glutaraldehyde sterilizing solutions. Infect Control. 1980;1:90–92. [PubMed] [Google Scholar]

104. Scott EM, Gorman SP. Glutaraldehyde. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 361–381. [Google Scholar]

105. Rutala WA, Gergen MF, Weber DJ. Inactivation of Clostridium difficile spores by disinfectants. Infect Control Hosp Epidemiol. 1993;14:36–39. [PubMed] [Google Scholar]

106. Dyas A, Das BC. The activity of glutaraldehyde against Clostridium difficile. J Hosp Infect. 1985;6:41–45. [PubMed] [Google Scholar]

107. van Klingeren B, Pullen W. Glutaraldehyde resistant mycobacteria from endoscope washers. J Hosp Infect. 1993;25:147–149. [PubMed] [Google Scholar]

108. Griffiths PA, Babb JR, Bradley CR. Glutaraldehyde-resistant Mycobacterium chelonae from endoscope washer disinfectors. J Appl Microbiol. 1997;82:519–526. [PubMed] [Google Scholar]

109. Fisher CW, Fiorello A, Shaffer D. Aldehyde-resistant mycobacteria bacteria associated with the use of endoscope reprocessing systems. Am J Infect Control. 2012;40:880–882. [PMC free article] [PubMed] [Google Scholar]

110. Webster E, Ribner B, Streed LL. Microbial contamination of activated 2% glutaraldehyde used in high-level disinfection of endoscopes [abstract] Am J Infect Control. 1996;24:153. [Google Scholar]

111. Barbee SL, Weber DJ, Sobsey MD, Rutala WA. Inactivation of Cryptosporidium parvum oocyst infectivity by disinfection and sterilization processes. Gastrointest Endosc. 1999;49:605–611. [PubMed] [Google Scholar]

112. Laskowski LF, Marr JJ, Spernoga JF. Fastidious mycobacteria grown from porcine prosthetic-heart-valve cultures. N Engl J Med. 1977;297:101–102. [PubMed] [Google Scholar]

113. Lorena NS, Duarte RS, Pitombo MB. Vol. 36. 2009. Rapidly growing mycobacteria infection after videosurgical procedures-the glutaraldehyde hypothesis; pp. 266–267. (Rev Col Bras Cir). [PubMed] [Google Scholar]

114. Svetlikova Z, Skovierova H, Niederweis M. Role of porins in the susceptibility of Mycobacterium smegmatis and Mycobacterium chelonae to aldehyde-based disinfectants and drugs. Antimicrob Agents Chemother. 2009;53:4015–4018. [PMC free article] [PubMed] [Google Scholar]

115. Mbithi JN, Springthorpe VS, Sattar SA. Bactericidal, virucidal, and mycobactericidal activities of reused alkaline glutaraldehyde in an endoscopy unit. J Clin Microbiol. 1993;31:2988–2995. [PMC free article] [PubMed] [Google Scholar]

116. Cole EC, Rutala WA, Nessen L. Effect of methodology, dilution, and exposure time on the tuberculocidal activity of glutaraldehyde-based disinfectants. Appl Environ Microbiol. 1990;56:1813–1817. [PMC free article] [PubMed] [Google Scholar]

117. Collins FM, Montalbine V. Mycobactericidal activity of glutaraldehyde solutions. J Clin Microbiol. 1976;4:408–412. [PMC free article] [PubMed] [Google Scholar]

118. Overton D, Burgess JO, Beck B. Glutaraldehyde test kits: evaluation for accuracy and range. Gen Dent. 1989;37:126. [PubMed] [Google Scholar]

119. Cooke RPD, Goddard SV, Chatterley R. Monitoring glutaraldehyde dilution in automated washer/disinfectors. J Hosp Infect. 2001;48:242–246. [PubMed] [Google Scholar]

120. Farina A, Fievet MH, Plassart F. Residual glutaraldehyde levels in fiberoptic endoscopes: measurement and implications for patient toxicity. J Hosp Infect. 1999;43:293–297. [PubMed] [Google Scholar]

121. Dailey JR, Parnes RE, Aminlari A. Glutaraldehyde keratopathy. Am J Ophthalmol. 1993;115:256–258. [PubMed] [Google Scholar]

122. Newman MA, Kachuba JB. Glutaraldehyde: a potential health risk to nurses. Gastroenterol Nurs. 1992;14:296–300. [PubMed] [Google Scholar]

123. Turner FJ. Hydrogen peroxide and other oxidant disinfectants. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 3rd ed. Lea & Febiger; Philadelphia: 1983. pp. 240–250. [Google Scholar]

124. Block SS. Peroxygen compounds. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 185–204. [Google Scholar]

125. Sattar SA, Springthorpe VS, Rochon M. A product based on accelerated and stabilized hydrogen peroxide: evidence for broad-spectrum germicidal activity. Can J Infect Control. 1998;(Winter):123–130. [Google Scholar]

126. Rutala WA, Gergen MF, Weber DJ. Sporicidal activity of chemical sterilants used in hospitals. Infect Control Hosp Epidemiol. 1993;14:713–718. [PubMed] [Google Scholar]

127. Maillard J-Y. Innate resistance to sporicides and potential failure to decontaminate. J Hosp Infect. 2011;77:204–209. [PubMed] [Google Scholar]

128. Silvany RE, Dougherty JM, McCulley JP. The effect of currently available contact lens disinfection systems on Acanthamoeba castellanii and Acanthamoeba polyphaga. Ophthalmology. 1990;97:286–290. [PubMed] [Google Scholar]

129. Neely AN, Maley MP. The 1999 Lindberg award. 3% hydrogen peroxide for the gram-positive disinfection of fabrics. J Burn Care Rehabil. 1999;20:471–477. [PubMed] [Google Scholar]

130. Vesley D, Norlien KG, Nelson B. Significant factors in the disinfection and sterilization of flexible endoscopes. Am J Infect Control. 1992;20:291–300. [PubMed] [Google Scholar]

131. Levenson JE. Corneal damage from improperly cleaned tonometer tips. Arch Ophthalmol. 1989;107:1117. [PubMed] [Google Scholar]

132. Omidbakhsh N, Sattar SA. Broad-spectrum microbicidal activity, toxicologic assessment, and materials compatibility of a new generation of accelerated hydrogen peroxide-based environmental surface disinfectant. Am J Infect Control. 2006;34:251–257. [PMC free article] [PubMed] [Google Scholar]

133. Sattar SA, Adegbunrin O, Ramirez J. Combined application of simulated reuse and quantitative carrier test to assess high-level disinfection: experiments with an accelerated hydrogen peroxide-based formulation. Am J Infect Control. 2002;30:449–457. [PubMed] [Google Scholar]

134. Omidbakhsh N. A new peroxide-based flexible endoscope-compatible high-level disinfectant. Am J Infect Control. 2006;34:571–577. [PubMed] [Google Scholar]

135. Omidbakhsh N, Kenny N. An accelerated hydrogen peroxide (AHP)-based fast-acting and reusable microbicide for manual disinfection of heat sensitive semi-critical medical devices. Can J Infect Control. 2008;23:81–88. [PubMed] [Google Scholar]

136. Rochon M, Sullivan N. Products based on accelerated and stabilized hydrogen peroxide: evidence for cleaning and sanitizing efficiency, environmental and human safety and non-corrosiveness. Can J Infect Control. 1999;(Summer):51–55. [Google Scholar]

137. Gottardi W. Iodine and iodine compounds. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 159–184. [Google Scholar]

138. Craven DE, Moody B, Connolly MG. Pseudobacteremia caused by povidone-iodine solution contaminated with Pseudomonas cepacia. N Engl J Med. 1981;305:621–623. [PubMed] [Google Scholar]

139. Berkelman RL, Lewin S, Allen JR. Pseudobacteremia attributed to contamination of povidone-iodine with Pseudomonas cepacia. Ann Intern Med. 1981;95:32–36. [PubMed] [Google Scholar]

140. Weber DJ, Rutala WA, Sickbert-Bennett EE. Outbreaks associated with contaminated antiseptics and disinfectants. Antimicrob Agents Chemother. 2007;51:916–919. [PMC free article] [PubMed] [Google Scholar]

141. Rutala WA, Cole EC, Wannamaker NS. Inactivation of Mycobacterium tuberculosis and Mycobacterium bovis by 14 hospital disinfectants. Am J Med. 1991;91:267S–271S. [PubMed] [Google Scholar]

142. Klein M, DeForest A. Vol. 49. 1963. The inactivation of viruses by ger­micides; pp. 116–118. (Chem Specialists Manuf Assoc Proc). [Google Scholar]

143. Berkelman RL, Holland BW, Anderson RL. Increased bactericidal activity of dilute preparations of povidone-iodine solutions. J Clin Microbiol. 1982;15:635–639. [PMC free article] [PubMed] [Google Scholar]

144. Wallbank AM, Drulak M, Poffenroth L. Wescodyne: lack of activity against poliovirus in the presence of organic matter. Health Lab Sci. 1978;15:133–137. [PubMed] [Google Scholar]

145. Favero MS, Bond WW. Chemical disinfection of medical and surgical materials. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 4th ed. Lea & Febiger; Philadelphia: 1991. pp. 617–641. [Google Scholar]

147. Gordon MD, Ezzell RJ, Bruckner NI. Enhancement of mycobactericidal activity of glutaraldehyde with α,β-unsaturated and aromatic aldehydes. J Indus Microbiol. 1994;13:77–82. [Google Scholar]

148. Rutala WA, Weber DJ. New disinfection and sterilization methods. Emerg Infect Dis. 2001;7:348–353. [PMC free article] [PubMed] [Google Scholar]

149. Gregory AW, Schaalje GB, Smart JD. The mycobactericidal efficacy of ortho-phthalaldehyde and the comparative resistances of Mycobacterium bovis, Mycobacterium terrae, and Mycobacterium chelonae. Infect Control Hosp Epidemiol. 1999;20:324–330. [PubMed] [Google Scholar]

150. Walsh SE, Maillard JY, Russell AD. Ortho-phthalaldehyde: a possible alternative to glutaraldehyde for high level disinfection. J Appl Microbiol. 1999;86:1039–1046. [PubMed] [Google Scholar]

151. Alfa MJ, Sitter DL. In-hospital evaluation of ortho-phthalaldehyde as a high level disinfectant for flexible endoscopes. J Hosp Infect. 1994;26:15–26. [PMC free article] [PubMed] [Google Scholar]

152. Roberts CG, Chan-Myers HB, Favero MS. Virucidal activity of ortho-phthalaldehyde solutions against hepatitis B and C viruses. Am J Infect Control. 2008;36:223–226. [PubMed] [Google Scholar]

153. Sokol WN. Nine episodes of anaphylaxis following cystoscopy caused by Cidex OPA (ortho-phthalaldehyde) high-level disinfectant in 4 patients after cystoscopy. J Allergy Clin Immunol. 2004;114:392–397. [PubMed] [Google Scholar]

154. Miyajima K, Yoshida J, Kumagai S. Ortho-phthalaldehyde exposure levels among endoscope disinfection workers. Sangyo Eiseigaku Zasshi. 2010;52:74. [PubMed] [Google Scholar]

155. Tucker RC, Lestini BJ, Marchant RE. Surface analysis of clinically used expanded PTFE endoscopic tubing treated by the STERIS PROCESS. ASAIO J. 1996;42:306–313. [PubMed] [Google Scholar]

157. Malchesky PS. Medical applications of peracetic acid. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 979–996. [Google Scholar]

158. Mannion PT. The use of peracetic acid for the reprocessing of flexible endoscopes and rigid cystoscopes and laparoscopes. J Hosp Infect. 1995;29:313–315. [PubMed] [Google Scholar]

159. Bradley CR, Babb JR, Ayliffe GA. Evaluation of the Steris System 1 Peracetic Acid Endoscope Processor. J Hosp Infect. 1995;29:143–151. [PubMed] [Google Scholar]

160. Duc DL, Ribiollet A, Dode X. Evaluation of the microbicidal efficacy of Steris System 1 for digestive endoscopes using GERMANDE and ASTM validation protocols. J Hosp Infect. 2001;48:135–141. [PubMed] [Google Scholar]

161. Alfa MJ, Olson N, Degagne P. New low temperature sterilization technologies: microbicidal activity and clinical efficacy. In: Rutala WA, editor. Disinfection, Sterilization, and Antisepsis in Healthcare. Polyscience Publications; Champlain, NY: 1998. pp. 67–78. [Google Scholar]

162. Alfa MJ, DeGagne P, Olson N. Comparison of liquid chemical sterilization with peracetic acid and ethylene oxide sterilization for long narrow lumens. Am J Infect Control. 1998;26:469–477. [PubMed] [Google Scholar]

163. Fuselier HA, Jr, Mason C. Liquid sterilization versus high level disinfection in the urologic office. Urology. 1997;50:337–340. [PubMed] [Google Scholar]

164. Seballos RJ, Walsh AL, Mehta AC. Clinical evaluation of a liquid chemical sterilization system for flexible bronchoscopes. J Bronch. 1995;2:192–199. [Google Scholar]

165. Wallace CG, Agee PM, Demicco DD. Liquid chemical sterilization using peracetic acid: an alternative approach to endoscope processing. ASAIO J. 1995;41:151–154. [PubMed] [Google Scholar]

166. Centers for Disease Control and Prevention Bronchoscopy-related infections and pseudoinfections—New York, 1996 and 1998. MMWR Morb Mortal Wkly Rep. 1999;48:557–560. [PubMed] [Google Scholar]

167. U.S. Food and Drug Administration, Centers for Disease Control and Prevention . U.S. Food and Drug Administration; Rockville, MD: 1999. FDA and CDC Public Health Advisory: Infections from Endoscopes Inadequately Reprocessed by an Automated Endoscope Reprocessing System. [Google Scholar]

168. Middleton AM, Chadwick MV, Gaya H. Disinfection of bronchoscopes, contaminated in vitro with Mycobacterium tuberculosis, Mycobacterium avium-intracellulare and Mycobacterium chelonae in sputum, using stabilized, buffered peracetic acid solution (‘Nu-Cidex. J Hosp Infect. 1997;37:137–143. [PubMed] [Google Scholar]

169. Holton J, Shetty N. In-use stability of Nu-Cidex. J Hosp Infect. 1997;35:245–248. [PubMed] [Google Scholar]

170. Alasri A, Roques C, Michel G. Bactericidal properties of peracetic acid and hydrogen peroxide, alone and in combination, and chlorine and formaldehyde against bacterial water strains. Can J Microbiol. 1992;38:635–642. [PubMed] [Google Scholar]

171. Stanley P. Destruction of a glutaraldehyde-resistant mycobacterium by a per-oxygen disinfectant [abstract] Am J Infect Control. 1998;26:185. [Google Scholar]

172. Fleming SJ, Foreman K, Shanley K. Dialyser reprocessing with Renalin. Am J Nephrol. 1991;11:27–31. [PubMed] [Google Scholar]

173. Finelli L, Miller JT, Tokars JI. National surveillance of dialysis-associated diseases in the United States, 2002. Semin Dial. 2005;18:52–61. [PubMed] [Google Scholar]

174. Rutala WA, Cole EC. Ineffectiveness of hospital disinfectants against bacteria: a collaborative study. Infect Control. 1987;8:501–506. [PubMed] [Google Scholar]

175. Prindle RF. Phenolic compounds. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 3rd ed. Lea & Febiger; Philadelphia: 1983. pp. 197–224. [Google Scholar]

176. Cole EC, Rutala WA, Samsa GP. Disinfectant testing using a modified use-dilution method: collaborative study. J Assoc Off Anal Chem. 1988;71:1187–1194. [PubMed] [Google Scholar]

177. Goddard PA, McCue KA. Phenolic compounds. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 255–281. [Google Scholar]

178. Sagripanti JL, Eklund CA, Trost PA. Comparative sensitivity of 13 species of pathogenic bacteria to seven chemical germicides. Am J Infect Control. 1997;25:335–339. [PubMed] [Google Scholar]

179. Wysowski DK, Flynt JW, Jr, Goldfield M. Epidemic neonatal hyperbilirubinemia and use of a phenolic disinfectant detergent. Pediatrics. 1978;61:165–170. [PubMed] [Google Scholar]

180. Doan HM, Keith L, Shennan AT. Phenol and neonatal jaundice. Pediatrics. 1979;64:324–325. [PubMed] [Google Scholar]

181. Shickman MD, Guze LB, Pearce ML. Bacteremia following cardiac catheterization. N Engl J Med. 1959;260:1164–1166. [PubMed] [Google Scholar]

182. Ehrenkranz NJ, Bolyard EA, Wiener M. Antibiotic-sensitive Serratia marcescens infections complicating cardiopulmonary operations: contaminated disinfectant as a reservoir. Lancet. 1980;2:1289–1292. [PubMed] [Google Scholar]

183. Mbithi JN, Springthorpe VS, Sattar SA. Chemical disinfection of hepatitis A virus on environmental surfaces. Appl Environ Microbiol. 1990;56:3601–3604. [PMC free article] [PubMed] [Google Scholar]

184. Petrocci AN. Surface active agents: quaternary ammonium compounds. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 3rd ed. Lea & Febiger; Philadelphia: 1983. pp. 309–329. [Google Scholar]

185. Doultree JC, Druce JD, Birch CJ. Inactivation of feline calicivirus, a Norwalk virus surrogate. J Hosp Infect. 1999;41:51–57. [PubMed] [Google Scholar]

186. Cefai C, Richards J, Gould FK. An outbreak of respiratory tract infection resulting from incomplete disinfection of ventilatory equipment. J Hosp Infect. 1990;15:177–182. [PubMed] [Google Scholar]

187. Gurevich I, Tafuro P, Ristuccia P. Disinfection of respirator tubing: a comparison of chemical versus hot water machine-assisted processing. J Hosp Infect. 1983;4:199–208. [PubMed] [Google Scholar]

188. Rutala WA, Weber DJ, Gergen MF. Efficacy of a washer-pasteurizer for disinfection of respiratory-care equipment. Infect Control Hosp Epidemiol. 2000;21:333–336. [PubMed] [Google Scholar]

189. Jette LP, Lambert NG. Evaluation of two hot water washer disinfectors for medical instruments. Infect Control Hosp Epidemiol. 1988;9:194–199. [PubMed] [Google Scholar]

190. Bak J, Ladefoged SD, Tvede M. Dose requirements for UVC disinfection of catheter biofilms. Biofouling. 2009;25:289–296. [PubMed] [Google Scholar]

191. Kac G, Gueneret M, Rodi A. Evaluation of a new disinfection procedure for ultrasound probes using ultraviolet light. J Hosp Infect. 2007;65:163–168. [PubMed] [Google Scholar]

192. Rastogi VK, Wallace L, Smith LS. Disinfection of Acinetobacter baumannii–contaminated surfaces relevant to medical treatment facilities with ultraviolet C light. Mil Med. 2007;172:1166–1169. [PubMed] [Google Scholar]

193. Nerandzic MM, Cadnum JL, Eckart KE. Evaluation of a hand-held far-ultraviolet radiation device for decontamination of Clostridium difficile and other healthcare-associated pathogens. BMC Infect Dis. 2012;12:120. [PMC free article] [PubMed] [Google Scholar]

194. Association for the Advancement of Medical Instrumentation, American National Standards Institute . Association for the Advancement of Medical Instrumentation; Arlington, VA: 2010. Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities. [Google Scholar]

195. Singh J, Bhatia R, Gandhi JC. Outbreak of viral hepatitis B in a rural community in India linked to inadequately sterilized needles and syringes. Bull World Health Organ. 1998;76:93–98. [PMC free article] [PubMed] [Google Scholar]

196. Eickhoff TC. An outbreak of surgical wound infections due to Clostridium perfringens. Surg Gynecol Obstet. 1962;114:102–108. [PubMed] [Google Scholar]

197. Tosh PK, Disbot M, Duffy JM. Outbreak of Pseudomonas aeruginosa surgical site infections after arthroscopic procedures: Texas, 2009. Infect Control Hosp Epidemiol. 2011;32:1179–1186. [PubMed] [Google Scholar]

198. Favero MS. Sterility assurance: concepts for patient safety. In: Rutala WA, editor. Disinfection, Sterilization and Antisepsis: Principles and Practices in Healthcare Facilities. Association for Professional in Infection Control and Epidemiology; Washington, DC: 2001. pp. 110–119. [Google Scholar]

199. Oxborrow GS, Berube R. Sterility testing-validation of sterilization processes, and sporicide testing. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 4th ed. Lea & Febiger; Philadelphia: 1991. pp. 1047–1057. [Google Scholar]

200. Rutala WA, Weber DJ. Clinical effectiveness of low-temperature sterilization technologies. Infect Control Hosp Epidemiol. 1998;19:798–804. [PubMed] [Google Scholar]

201. Adler S, Scherrer M, Daschner FD. Costs of low-temperature plasma sterilization compared with other sterilization methods. J Hosp Infect. 1998;40:125–134. [PubMed] [Google Scholar]

202. Joslyn L. Sterilization by heat. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 695–728. [Google Scholar]

203. Silverstone SE, Hill DE. Evaluation of sterilization of dental handpieces by heating in synthetic compressor lubricant. Gen Dent. 1999;47:158–160. [PubMed] [Google Scholar]

204. Bucx MJ, Veldman DJ, Beenhakker MM. The effect of steam sterilization at 134 degrees C on light intensity provided by fibrelight Macintoch laryngoscopes. Anaesthesia. 2000;55:185–186. [PubMed] [Google Scholar]

205. Gilbert JA, Phillips HO. The effect of steam sterilization on plaster casting material. Clin Orthop Relat Res. 1984;(190):241–244. [PubMed] [Google Scholar]

206. Bryce EA, Roberts FJ, Clements B. When the biological indicator is positive: investigating autoclave failures. Infect Control Hosp Epidemiol. 1997;18:654–656. [PubMed] [Google Scholar]

207. Rutala WA, Stiegel MM, Sarubbi FA., Jr Decontamination of laboratory microbiological waste by steam sterilization. Appl Environ Microbiol. 1982;43:1311–1316. [PMC free article] [PubMed] [Google Scholar]

208. Rutala WA. Disinfection and flash sterilization in the operating room. J Ophthalm Nurs Technol. 1991;10:106–115. [PubMed] [Google Scholar]

209. Seavey R. Immediate use steam sterilization: moving beyond current policy. Am J Infect Control. 2013;41(5 suppl):S46–S48. [PubMed] [Google Scholar]

211. Ernst RR, Doyle JE. Sterilization with gaseous ethylene oxide: a review of chemical and physical factors. Biotech Bioeng. 1968;10:1–31. [Google Scholar]

212. Joslyn L. Gaseous chemical sterilization. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 337–360. [Google Scholar]

213. Association for the Advancement of Medical Instrumentation . AAMI; Arlington, VA: 1999. Ethylene Oxide Sterilization in Health Care Facilities: Safety and Effectiveness. Report No. ANSI/AAMI ST41. [Google Scholar]

214. Association for the Advancement of Medical Instrumentation . AAMI; Arlington, VA: 2008. Ethylene Oxide Sterilization in Health Care Facilities: Safety and Effectiveness. Report No. ANSI/AAMI ST41. [Google Scholar]

216. Ethylene oxide sterilization: how hospitals can adapt to the changes. Health Devices. 1994;23:485–492. [PubMed] [Google Scholar]

217. Ries MD, Weaver K, Beals N. Safety and efficacy of ethylene oxide sterilized polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 1996;(331):159–163. [PubMed] [Google Scholar]

218. Alfa MJ, DeGagne P, Olson N. Bacterial killing ability of 10% ethylene oxide plus 90% hydrochlorofluorocarbon sterilizing gas. Infect Control Hosp Epidemiol. 1997;18:641–645. [PubMed] [Google Scholar]

219. Rutala WA, Gergen MF, Weber DJ. Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. Am J Infect Control. 1998;26:393–398. [PubMed] [Google Scholar]

220. Parisi AN, Young WE. Sterilization with ethylene oxide and other gases. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 4th ed. Lea & Febiger; Philadelphia: 1991. pp. 580–595. [Google Scholar]

221. Holler C, Martiny H, Christiansen B. The efficacy of low temperature plasma (LTP) sterilization, a new sterilization technique. Zentralbl Hyg Umweltmed. 1993;194:380–391. [PubMed] [Google Scholar]

222. Parker HH, Johnson RB. Effectiveness of ethylene oxide for sterilization of dental handpieces. J Dent. 1995;23:113–115. [PubMed] [Google Scholar]

223. Borneff M, Ruppert J, Okpara J. Efficacy testing of low-temperature plasma sterilization (LTP) with test object models simulating practice conditions. Zentralbl Steril. 1995;3:361–371. [Google Scholar]

224. Borneff-Lipp M, Okpara J, Bodendorf M. Validation of low-temperature-plasma (LPT) sterilization systems: comparison of two technical versions, the Sterrad 100, 1.8 and the 100S. Hyg Mikrobiol. 1997;3:21–28. [Google Scholar]

225. Jacobs PT, Lin SM. Sterilization processes utilizing low-temperature plasma. In: Block SS, editor. Disinfection, Sterilization, and Preservation. 5th ed. Lippincott Williams & Wilkins; Philadelphia: 2001. pp. 747–763. [Google Scholar]

226. Rutala WA, Gergen MF, Weber DJ. Sporicidal activity of a new low-temperature sterilization technology: the Sterrad 50 sterilizer. Infect Control Hosp Epidemiol. 1999;20:514–516. [PubMed] [Google Scholar]

227. Roberts C, Antonoplos P. Inactivation of human immunodeficiency virus type 1, hepatitis A virus, respiratory syncytial virus, vaccinia virus, herpes simplex virus type 1, and poliovirus type 2 by hydrogen peroxide gas plasma sterilization. Am J Infect Control. 1998;26:94–101. [PubMed] [Google Scholar]

228. Kyi MS, Holton J, Ridgway GL. Assessment of the efficacy of a low temperature hydrogen peroxide gas plasma sterilization system. J Hosp Infect. 1995;31:275–284. [PubMed] [Google Scholar]

229. Borneff-Lipp M, Kaetzke A, Durr M. Evaluation of low temperature hydrogen peroxide plasma sterilization. Zentralbl Steril. 2008;16:35–42. [Google Scholar]

230. Feldman LA, Hui HK. Compatibility of medical devices and materials with low-temperature hydrogen peroxide gas plasma. Med Dev Diag Indust. 1997;19:57–62. [Google Scholar]

231. Jacobs PT, Smith D. The new Sterrad 100S sterilization system: features and advantages. Zentralbl Steril. 1998;6:86–94. [Google Scholar]

232. Schembre DB. Infectious complications associated with gastrointestinal endoscopy. Gastrointest Endosc Clin North Am. 2000;10:215–232. [PubMed] [Google Scholar]

233. Weber DJ, Rutala WA. Lessons learned from outbreaks and pseudo-outbreaks associated with bronchoscopy. Infect Control Hosp Epidemiol. 2012;33:224–229. [PubMed] [Google Scholar]

234. Nelson DB, Jarvis WR, Rutala WA. Multi-society guideline for reprocessing flexible gastrointestinal endoscopes. Infect Control Hosp Epidemiol. 2003;24:532–537. [PubMed] [Google Scholar]

235. Society of Gastroenterology Nurses and Associates SGNA Standards of infection control in reprocessing of flexible gastrointestinal endoscopes. Gastroenterol Nurs. 2010;33:70–80. [PubMed] [Google Scholar]

236. Ofstead CL, Wetzler HP, Snyder AK. Endoscope reprocessing methods: a prospective study on the impact of human factors and automation. Gastroenterol Nurs. 2010;33:304–311. [PubMed] [Google Scholar]

237. Jackson FW, Ball MD. Correction of deficiencies in flexible fiberoptic sigmoidoscope cleaning and disinfection technique in family practice and internal medicine offices. Arch Fam Med. 1997;6:578–582. [PubMed] [Google Scholar]

238. Orsi GB, Filocamo A, Di Stefano L. Italian National Survey of Digestive Endoscopy Disinfection Procedures. Endoscopy. 1997;29:732–738. [PubMed] [Google Scholar]

239. Honeybourne D, Neumann CS. An audit of bronchoscopy practice in the United Kingdom: a survey of adherence to national guidelines. Thorax. 1997;52:709–713. [PMC free article] [PubMed] [Google Scholar]

240. Srinivasan A, Wolfenden LL, Song X. An outbreak of Pseudomonas aeruginosa infections associated with flexible bronchoscopes. N Engl J Med. 2003;348:221–227. [PubMed] [Google Scholar]

241. Cetse JC, Vanhems P. Outbreak of infection associated with bronchoscopes. N Engl J Med. 2003;348:2039–2040. [PubMed] [Google Scholar]

242. Rutala WA, Weber DJ. How to assess risk of disease transmission when there is a failure to follow recommended disinfection and sterilization principles. Infect Control Hosp Epidemiol. 2007;28:519–524. [PubMed] [Google Scholar]

243. Rutala WA, Weber DJ. FDA labeling requirements for disinfection of endoscopes: a counterpoint. Infect Control Hosp Epidemiol. 1995;16:231–235. [PubMed] [Google Scholar]

244. Kruse A, Rey JF. Guidelines on cleaning and disinfection in GI endoscopy: update 1999. The European Society of Gastrointestinal Endoscopy. Endoscopy. 2000;32:77–80. [PubMed] [Google Scholar]

245. British Society of Gastroenterology Cleaning and disinfection of equipment for gastrointestinal endoscopy. Report of a working party of the British Society of Gastroenterology Endoscope Committee. Gut. 1998;42:585–593. [PMC free article] [PubMed] [Google Scholar]

246. British Thoracic Society British Thoracic Society guidelines on diagnostic flexible bronchoscopy. Thorax. 2001;56(suppl 1):1–21. [PMC free article] [PubMed] [Google Scholar]

247. Occupational Safety and Health Administration Occupational exposure to bloodborne pathogens; final rule. Fed Register. 1991;56:64003–64182. [Google Scholar]

248. Occupational Safety and Health Administration . 1992. OSHA Instruction CPL 2-2.44C. Office of Health Compliance Assistance. Washington, DC, March 6. [Google Scholar]

249. Occupational Safety and Health Administration . 1997. OSHA Memorandum from Stephen Mallinger. EPA-Registered Disinfectants for HIV/HBV. Washington, DC; February 28. [Google Scholar]

250. Payan C, Cottin J, Lemarie C. Inactivation of hepatitis B virus in plasma by hospital in-use chemical disinfectants assessed by a modified HepG2 cell culture. J Hosp Infect. 2001;47:282–287. [PubMed] [Google Scholar]

251. Henderson DA. The looming threat of bioterrorism. Science. 1999;283:1279–1282. [PubMed] [Google Scholar]

252. Centers for Disease Control and Prevention Biological and chemical terrorism: strategic plan for preparedness and response. MMWR Recomm Rep. 2000;49(RR-4):1–14. [PubMed] [Google Scholar]

253. Rutala WA, Weber DJ. Infection control: the role of dis­infection and sterilization. J Hosp Infect. 1999;43(suppl):S43–S55. [PubMed] [Google Scholar]

254. Ryan SP, Calfee MW, Wood JP. Research to support the decontamination of surfaces and buildings contaminated with biothreat agents. In: Rutala WA, editor. Disinfection, Sterilization and Antisepsis: Principles, Practices, Current Issues, New Research, and New Technologies. Association for Professionals in Infection Control and Epidemiology; Washington, DC: 2010. pp. 260–306. [Google Scholar]

255. Rutala WA, Stiegel MM, Sarubbi FA. Susceptibility of antibiotic-susceptible and antibiotic-resistant hospital bacteria to disinfectants. Infect Control Hosp Epidemiol. 1997;18:417–421. [PubMed] [Google Scholar]

256. Weber DJ, Rutala WA. Use of germicides in the home and health care setting: is there a relationship between germicide use and antimicrobial resistance. Infect Control Hosp Epidemiol. 2006;27:1107–1119. [PubMed] [Google Scholar]

257. Rutala WA, Weber DJ. Guideline for disinfection and sterilization of prion-contaminated medical instruments. Infect Control Hosp Epidemiol. 2010;31:107–117. [PubMed] [Google Scholar]

258. Centers for Disease Control and Prevention Surveillance for Creutzfeldt-Jakob disease—United States. MMWR Morb Mortal Wkly Rep. 1996;45:665–668. [PubMed] [Google Scholar]

259. Belay ED, Schonberger LB, Brown P. Disinfection and sterilization of prion-contaminated medical instruments. Infect Control Hosp Epidemiol. 2010;31:1304–1306. [PubMed] [Google Scholar]

260. Rutala WA, Weber DJ. Disinfection and sterilization of prion-contaminated medical instruments: reply to Belay. Infect Control Hosp Epidemiol. 2010;31:1306–1307. [Google Scholar]

261. Secker TJ, Herve R, Keevil CW. Adsorption of prion and tissue proteins to surgical stainless steel surfaces and the efficacy of decontamination following dry and wet storage conditions. J Hosp Infect. 2011;78:251–255. [PubMed] [Google Scholar]

263. Rutala WA, Weber DJ. Creutzfeldt-Jakob disease: recommendations for disinfection and sterilization. Clin Infect Dis. 2001;32:1348–1356. [PubMed] [Google Scholar]

264. Taylor DM. Inactivation of prions by physical and chemical means. J Hosp Infect. 1999;43(suppl):S69–S76. [PubMed] [Google Scholar]

265. Otter JA. The role played by contaminated surfaces in the transmission of nosocomial pathogens. Infect Control Hosp Epidemiol. 2011;32:687–699. [PubMed] [Google Scholar]

266. Boyce JM. Environmental contamination makes an important contribution to hospital infection. J Hosp Infect. 2007;65:50–54. [PubMed] [Google Scholar]

267. Weber DJ, Rutala WA, Miller MB. Role of hospital surfaces in the transmission of emerging health care-associated pathogens: norovirus, Clostridium difficile, and Acinetobacter species. Am J Infect Control. 2010;38:S25–S33. [PubMed] [Google Scholar]

268. Rutala WA, Weber DJ. Are room decontamination units needed to prevent transmission of environmental pathogens? Infect Control Hosp Epidemiol. 2011;32:743–747. [PubMed] [Google Scholar]

269. Stiefel U, Cadnum JL, Eckstein BC. Contamination of hands with methicillin-resistant Staphylococcus aureus after contact with environmental surfaces and after contact with the skin of colonized patients. Infect Control Hosp Epidemiol. 2011;32:185–187. [PubMed] [Google Scholar]

270. Huang SS, Datta R, Platt R. Risk of acquiring antibiotic-resistant bacteria from prior room occupants. Arch Intern Med. 2006;166:1945–1951. [PubMed] [Google Scholar]

271. Shaughnessy MK, Micielli RL, DePestal DD. Evaluation of hospital room assignment and acquisition of Clostridium difficile infection. Infect Control Hosp Epidemiol. 2011;32:201–206. [PubMed] [Google Scholar]

272. Carling PC, Parry MF, Von Beheren SM, Healthcare Environmental Hygiene Study Group Identifying opportunities to enhance environmental cleaning in 23 acute care hospitals. Infect Control Hosp Epidemiol. 2008;29:1–7. [PubMed] [Google Scholar]

273. Carling PC, von Bheren S, Kim P. Intensive care unit environmental cleaning: an evaluation in sixteen hospitals using a novel assessment tool. J Hosp Infect. 2008;68:39–44. [PubMed] [Google Scholar]

274. Havill NL, Havill HL, Mangione E. Cleanliness of portable medical equipment disinfected by nursing staff. Am J Infect Control. 2011;39:602–604. [PubMed] [Google Scholar]

275. Hota B, Blom DW, Lyle EA. Intervention evaluation of environmental contamination by vancomycin-resistant enterococci: failure of personnel, product, or procedure? J Hosp Infect. 2009;71:123–131. [PubMed] [Google Scholar]

276. Goodman ER, Platt R, Bass R. Impact of an environmental cleaning intervention on the presence of methicillin-resistant Staphylococcus aureus and vancomycin-resistant enterococci on surfaces in intensive care unit rooms. Infect Control Hosp Epidemiol. 2008;29:593–599. [PMC free article] [PubMed] [Google Scholar]

277. Eckstein BC, Adams DA, Eckstein EC. Reduction of Clostridium difficile and vancomycin-resistant Enterococcus contamination of environmental surfaces after an intervention to improve cleaning methods. BMC Infect Dis. 2007;7:61. [PMC free article] [PubMed] [Google Scholar]

278. Rutala WA, Weber DJ. Sterilization, high-level disinfection, and environmental cleaning. Infect Dis Clin North Am. 2011;25:45–76. [PubMed] [Google Scholar]

279. Memarzadeh F, Olmsted RN, Bartley JM. Applications of ultraviolet germicidal irradiation disinfection in health care facilities: effective adjunct, but not stand-alone technology. Am J Infect Control. 2010;38:S13–S24. [PMC free article] [PubMed] [Google Scholar]

280. Rutala WA, Gergen MF, Weber DJ. Room decontamination by ultraviolet radiation. Infect Control Hosp Epidemiol. 2010;31:1025–1029. [PubMed] [Google Scholar]

281. Boyce JM, Havill NL, Moore BA. Terminal decontamination of patient rooms using an automated mobile UV light unit. Infect Control Hosp Epidemiol. 2011;32:743–747. [PubMed] [Google Scholar]

282. Nerandzic MM, Cadnum JL, Pultz MJ. Evaluation of an automated ultraviolet radiation device for decon­tamination of Clostridium difficile and other healthcare-associated pathogens in hospital rooms. BMC Infect Dis. 2010;10:197. [PMC free article] [PubMed] [Google Scholar]

283. Boyce JM, Havill NL, Otter JA. Impact of hydrogen peroxide vapor room decontamination on Clostridium difficile environmental contamination and transmission in a healthcare setting. Infect Control Hosp Epidemiol. 2008;29:723–729. [PubMed] [Google Scholar]

284. French GL, Otter JA, Shannon KP. Tackling contamination of the hospital environment by methicillin-resistant Staphylococcus aureus (MRSA): a comparison between conventional terminal cleaning and hydrogen peroxide vapour decontamination. J Hosp Infect. 2004;57:31–37. [PubMed] [Google Scholar]

285. Bartels MD, Kristofferson K, Slotsbjerg T. Environmental methicillin-resistant Staphylococcus aureus (MRSA) disinfection using dry-mist-generated hydrogen peroxide. J Hosp Infect. 2008;70:35–41. [PubMed] [Google Scholar]

286. Hall L, Otter JA, Chewins J. Use of hydrogen peroxide vapor for deactivation of Mycobacterium tuberculosis in a biological safety cabinet and a room. J Clin Microbiol. 2007;45:810–815. [PMC free article] [PubMed] [Google Scholar]

287. Hardy KJ, Gossain S, Henderson N. Rapid recontamination with MRSA of the environment of an intensive care unit after decontamination with hydrogen peroxide vapour. J Hosp Infect. 2007;66:360–368. [PubMed] [Google Scholar]

288. Johnston MD, Lawson S, Otter JA. Evaluation of hydrogen peroxide vapour as a method for the decontamination of surfaces contaminated with Clostridium botulinum spores. J Microbiol Methods. 2005;60:403–411. [PubMed] [Google Scholar]

289. Heckert RA, Best M, Jordan LT. Efficacy of vaporized hydrogen peroxide against exotic animal viruses. Appl Environ Microbiol. 1997;63:3916–3918. [PMC free article] [PubMed] [Google Scholar]

290. Klapes NA, Vesley D. Vapor-phase hydrogen peroxide as a surface decontaminant and sterilant. Appl Environ Microbiol. 1990;56:503–506. [PMC free article] [PubMed] [Google Scholar]

291. Bates CJ, Pearse R. Use of hydrogen peroxide vapour for environmental control during a Serratia outbreak in a neonatal intensive care unit. J Hosp Infect. 2005;61:364–366. [PubMed] [Google Scholar]

292. Shapey S, Machin K, Levi K. Activity of a dry mist hydrogen peroxide system against environmental Clostridium difficile contamination in elderly care wards. J Hosp Infect. 2008;70:136–141. [PubMed] [Google Scholar]

293. Rutala WA, Odette RL, Samsa GP. Management of infectious waste by United States hospitals. JAMA. 1989;262:1635–1640. [PubMed] [Google Scholar]

294. Rutala WA, Sarubbi FA. Management of infectious waste from hospitals. Infect Control. 1983;4:198–204. [PubMed] [Google Scholar]

295. Agency for Toxic Substances and Disease Registry . U.S. Department of Health and Human Services; Washington, DC: 1990. The Public Health Implications of Medical Waste: A Report to Congress. [Google Scholar]

296. Environmental Protection Agency . U.S. Environmental Protection Agency; Washington, DC: 1986. EPA Guide for Infectious Waste Management. [Google Scholar]

297. Office of Technology Assessment (OTA-0-459 . Government Printing Office; Washington, DC: 1990. Finding the Rx for Managing Medical Wastes. [Google Scholar]

298. Environmental Protection Agency Standards for the tracking and management of medical waste; interim final rule and request for comments. Fed Register. 1989;54:12326–12395. [Google Scholar]

299. Rutala WA, Weber DJ. Infectious waste: mismatch between science and policy. N Engl J Med. 1991;325:578–582. [PubMed] [Google Scholar]

300. Environmental Protection Agency Hospital/Medical/Infectious Waste Incinerators: Summary of requirements for revised or new section 111(d)/129 State plans following amendments to the emission guidelines. 2010. http://www.epa.gov/ttn/atw/129/hmiwi/rihmiwi.html Available at.

301. Rutala WA, Mayhall CG, Society for Hospital Epidemiology of America SHEA position paper: Medical waste. Infect Control Hosp Epidemiol. 1992;13:38–48. [PubMed] [Google Scholar]

302. U.S. Department of Transportation Infectious substances: final rule. Fed Register. 1995;60:48779–48787. [Google Scholar]

303. Rutala WA. Selection and use of disinfectants in healthcare. In: Mayhall CG, editor. Hospital Epidemiology and Infection Control. 2nd ed. Lippincott Williams & Wilkins; Philadelphia: 1999. pp. 1161–1187. [Google Scholar]

304. Rutala WA, Weber DJ. Sterilization and disinfection. In: Jarvis WR, editor. Bennett & Brachman's Hospital Infections. 5th ed. Wolters Kluwer/Lippincott Williams & Wilkins; Philadelphia: 2007. pp. 303–318. [Google Scholar]

What is the best method of sterilization of medical instrument?

For many medical devices, sterilization with ethylene oxide may be the only method that effectively sterilizes and does not damage the device during the sterilization process.

Which sterilization method is most often used in a medical office quizlet?

Although in medical offices steam under pressure is the most frequently used method of sterilization, the method depends on the nature of the material to be sterilized and the type of microorganisms to be destroyed.

What is the most common sterilization process?

High-pressure, saturated steam using an autoclave, or dry heat using an oven, are the most common and readily available methods used for sterilization. Remember: When instruments and equipment are sterilized by high- pressure steam (autoclaving), it is essential that steam reach all surfaces.

Which sterilization method is used in hospital?

Steam under pressure, dry heat, ethylene oxide (ETO) gas, hydrogen peroxide gas plasma, vaporized hydrogen peroxide, and liquid chemicals are the principal sterilizing agents used in health care facilities.